Appendix B: - Customized PC Output Protocol
This instrument supports customized PC output protocol. Under this mode: -
•
2 data separation types
•
7 control codes, and
•
22 different transaction data
are available from instrument.
B.1 Customized PC Output Protocol Setting Procedures
1.
At Weighing mode, go to F16 or F17 depends on the comport # to be set. Press
[Print/M+]
.
2.
Select PC, then set baud rate, parity and data length.
3.
Select Custom in Protocol page. Press
[Print/M+]
.
4.
Instrument displays one of the data separators.
5.
Press
[Unit]
to shift between data separator Comma (comma) and SemiCo (semi colon) and press
[Print/M+]
.
Notes
: -
•
Data separator is a symbol inserted between transaction data and is usually used by computer
program to separate various data.
•
No data separator is inserted between in front or after control commands.
•
Separator in front and after a control code (C1 ~ C6) will not be surpassed. For example: -
STXdata
,
data
,
dataETX
•
Exception of above is when control codes are in between 2 data. For example: -
data
,
CR LF data
6.
Instrument displays item number, then followed by the content page.
Note: -
Item number means the output sequence, e.g. Item 1 = the first content to output, Item 3 = the
third content to output.
7.
Press
[Unit]
or
[MR]
until the preferred content appears and press
[Print/M+]
. Refer to
Customized
PC Output Content Table
for details.
8.
Repeat step
6
and
7
to include other transaction data or control.
9.
To complete and save a customized output, select
End
and press
[Print/M+]
.
10.
Then select output time interval. Refer to paragraphs
5.5.1
for details.
11.
Press
[Print/M+]
to save.