8. Piece Count Mode
a.
Refer to
6.6
on how to select the desired weight unit.
b.
If a container is used, place it onto the platform and press
[TARE]
.
c.
Apply samples with the known quantity (sample size) on platform.
d.
Press and hold
[FUNC/SET]
unit numeric digits flashing then input the sample quantity through numeric
keys then press
[PRINT/M+]
.
e.
This instrument will calculate, store the average piece weight and confirm with 2 beeps. The quantity
is then displayed.
f.
Add to or remove from the platform, the corresponding quantity will be displayed automatically.
g.
To count different articles, repeat procedures listed above.
8.1 Auto Unit Piece Weight Enhancement Function
In order to obtain the best counting result and minimize sampling error, this instrument is equipped with Auto
Unit Piece Weight Enhancement Function.
This function will automatically be employed when Unit piece weight is determined by actual sample method.
Auto Unit Piece Weight Enhancement Function is built-in the Piece Count mode. This function starts
automatically when all of the below are met: -
a.
Unit piece weight is determined by actual sampling method.
b.
The quantity
added
to platter is more than 5 pieces but less than current quantity on scale.
c.
The total quantity on scale is less than 10000 pieces.
When all the above requirements are met, a new unit piece weight will be calculated and stored in memory
and confirmed by a "beep" sound.
8.2 Shift among Quantity, Average Piece Weight and Weight Info
1.
Repeat press and hold
[MR/UNIT]
to shift among quantity, average piece weight and weight info.
2.
Quantity display format = numeric numbers & PCS (e.g1000 PCS). Average piece weight display format
= numeric numbers & weight unit &/PCS (e.g. 499.960g/PCS).
3.
Weight display format (when Piece Count Function is in effect) = numeric numbers & weight unit.
8.3 Recall the Average Piece Weight before Powered Off
Press
[FUNC/SET]
, then
[MR/UNIT]
to recall the last average piece weight before instrument was powered off.