1. 

Proper use and application  

1.1. 

Application 

The delivery table LM-03 is intended for lying and sitting, for patients at all the stages of labour. This 
table comes equipped with the permanent and extended lying surfaces. The lying surface height, and 
angles  of  the  backrest  and  seat  sections,  Trendelenburg  and  reverse  Trendelenburg,  are  electrically 
adjusted. 

1.2. 

General requirements 

The  device  is  intended  for indoor  use.  Required  climatic  conditions:  temperature  from  +10  to  +40°C, 
permissible  ambient  temperature  fluctuation  within  8 h:  below  20°C,  relative  air  humidity:  from  30  to 
80%,  atmospheric  pressure:  from  700  to  1060 hPa.  The  device  should  be  used,  operated  and 
maintained acc. to the principles specified in this manual.  
 

 

NOTE! 

Using,  operating  and  servicing  of  the  device  in  contravention  of  this  instruction 
manual is prohibited and may lead to damage, sustained by the user, for which the 
manufacturer assumes no liability. 

Mounting  and  installing  any  equipment  that  is  different  from  that  offered  by  the 
manufacturer is permitted only on the basis of the manufacturer's written consent. 

1.3. 

User responsibilities 

The user – any  entity holding an object  as an owner, tenant, pledgee, lessee or any entity  having a 
different  right  to  the  object,  further  –  any  entity  using  the  device  himself  or  on  behalf  of  whom  the 
device is used. 

The user is obliged to ensure that the device is operated in compliance with its intended purpose and 
that it is operated under appropriate conditions, in a manner pursuant to this instruction manual. The 
user is obliged to take all necessary measures to ensure safe operation of the device and to reduce all 
safety threats to life and health of both user and patient, as well as third parties. Any device operation 
is permitted only if by authorised persons, suitably trained and acquainted with this instruction manual. 
The user must ensure that all device operators read, understand and observe this instruction manual.  

1.4. 

Technical data 

 

dimensional length of the delivery table 

 

 

85 in (2,165 mm) ± 0.8 in (20 mm) 

 

dimensional length of the delivery table in the armchair position 

70 in (1,780 mm) ± 0.8 in (20 mm) 

 

max. width (including the rails) 

 

 

 

38 in (965 mm) ± 0.2 in (5 mm) 

 

min. height of the lying surface frame 

 

 

23 in (580 mm) ± 0.4 in (10 mm) 

 

max. height of the lying surface frame  

 

 

39 in (980 mm) ± 0.4 in (10 mm) 

 

 

 

 

 

 

seat elevation angle 

 

 

 

 

30° ± 5° 

 

backrest elevation angle 

 

 

 

 

70° ± 5° 

 

Trendelenburg 

 

 

 

 

 

20° ± 2° 

 

reverse Trendelenburg 

 

 

 

 

20° ± 2° 

 

delivery table weight 

 

 

 

 

ca. 25 st (160 kg) 

 

permissible load 

 

 

 

ca.  40 st  (250 kg),  that  is  patient:  34 st  (215 kg),  and 

equipment: 6 st (35 kg) 

 

permissible load of the extended lying surface 

27 st (170 kg) 

 

power supply 

 

 

 

 

230 V~ (optionally 100V~, 120V~), 50/60Hz 

 

power consumption  

 

 

 

max. 320VA/230V 

 

protection class 

 

 

 

 

II  

 

application part type 

 

 

 

 

protection degree   

 

 

 

at least IP-X4 

 

life span 

 

 

 

 

 

10 years 

 

On special request, we can manufacture the device with different technical parameters, without 
compromising its safety. 

Summary of Contents for FREYA 3

Page 1: ...INSTRUCTION MANUAL Delivery table type LM 03 FREYA 3 Serial no Revision 2 00 Publication date 26 03 2015 FAMED YWIEC Sp z o o Annexes 2 3 4 Index of Spare Parts...

Page 2: ...e essential requirements contained in Annex I to the directive and Act on medical devices The compliance procedure has been performed acc to Annex VII to the directive Manufacturer FAMED YWIEC Sp z o...

Page 3: ...ranty null and void Before any repairs disconnect the device from the mains Safety notes Below is the warning Warning Follow the instructions for safe use This warning can be found on parts and mechan...

Page 4: ...ly exercise caution while adjusting the extended lying surface in relation to the permanent one as the distance between the permanent frame and the extended one is too small if the pressure knob is us...

Page 5: ...r underneath must be free of obstacles During the table movement avoid collisions Exercise caution while removing and mounting the permanent lying surface moulding Application of excessive force may l...

Page 6: ...8 ARMCHAIR POSITION OF THE DELIVERY TABLE CARDIAC CHAIR 18 3 9 QUICK LEVELLING OF THE THIGH AND BACKREST SECTIONS CPR FUNCTION 18 3 10 POSITION 0 18 3 11 EXTENDED LYING SURFACE FRAME POSITION ADJUSTME...

Page 7: ...used The user is obliged to ensure that the device is operated in compliance with its intended purpose and that it is operated under appropriate conditions in a manner pursuant to this instruction man...

Page 8: ...nel with the option of locking the respective electrical functions of the delivery table Control panel in the side rail Wired controller Foot controllers option 9 Backlight lamp from underneath the de...

Page 9: ...related to the structure technology and materials without compromising the technical utility parameters and safety of the device 1 7 Safety The device structure ensures safe operation and use provide...

Page 10: ...nvironment is described in which the medical device Delivery Table LM 03 should be operated The user should observe the guidelines and warnings included in the tables NOTE The use of accessories addit...

Page 11: ...ransport method and package type ensuring safe transport 2 2 Unpacking and first start up NOTE If the device is mounted by the service team of FAMED YWIEC Sp z o o then the user is exempt from perform...

Page 12: ...r the electronic table control system works performing all the functional table movements by use of all the available control elements foot controller and side panels Furthermore check whether the loc...

Page 13: ...elected functions so as to prevent the patient from their activation with the wired controller or the ones in the side rails The central control panel enables the control of the following delivery tab...

Page 14: ...at the same time inclination angle of the backrest and thigh sections AUTOCONTOUR Trendelenburg and reverse Trendelenburg angles The table controller without the central control panel or locking pane...

Page 15: ...with control panels on the outer side see Fig 2a whereas on the inner side with control panels see Fig 2b which enable the activation of the table functions by the patient and medical staff as describ...

Page 16: ...e Trendelenburg when you lower the lying surface to the min height or during the Trendelenburg position marking compliant with pt 8 2 3 3 Backrest section angular position adjustment with the CPR func...

Page 17: ...central panel are marked with the symbol Angular couch position can be changed only by medical personnel The Trendelenburg position is a rescue position NOTE Do not lay any objects on the plastic base...

Page 18: ...e patient s head and push the backrest section down So as to level the thigh section pull the lever under the shank section and push the thigh section down 3 10 Position 0 The position 0 of the delive...

Page 19: ...e between the mounting element of the hand holder and the side rail Failure to observe the above note may at the extreme angles lead to crushing the patient s operator s fingers or to damaging the del...

Page 20: ...r underneath must be free of obstacles During the table movement avoid collisions Do not roll the delivery table across electrical cables 3 15 Use of the side rails Fig 4 The side rails fig 4 are atta...

Page 21: ...ht be damaged Application of excessive force might result in damaging the mouldings 3 17 Infusion rod The infusion rod is embedded in bushings located in the permanent lying surface frame near the hea...

Page 22: ...ually with the agents recommended in annex 2 to this instruction manual NOTE Use foil underlays for spots exposed to preparations applied during various procedures The contact with preparations for di...

Page 23: ...Instruction Manual for the Delivery Table FREYA 3 Page 23 4 Utility parameters The schematic figure below presents the most important utility functions of the delivery table and their marking...

Page 24: ...tion contained herein do not operate it Consult with the manufacturer supplier dealer Using the defective device may lead to damage sustained by the user for which the manufacturer assumes no liabilit...

Page 25: ...rotection NOTE The device contains products which may pose a threat to the environment gel battery type EP 1 2 12 2 pcs Rules of conduct as regards worn products that may pose a threat to the environm...

Page 26: ...should be settled acc to pt 6 6 Troubleshooting So as to ensure safe and reliable functioning within the device lifetime the user shall perform inspections as regards the compliance with standard IEC...

Page 27: ...pply cable disconnected controller defective control system or actuator connect the controller notify an authorised service centre seat section angle adjustment actuator will not work disconnected pow...

Page 28: ...ng features The user is to ensure that the medical device Delivery Table LM 03 is used in this type of environment Resistance IEC 60601 1 2 Test level Resistance level Explanations and guidelines Elec...

Page 29: ...a should be lower than the column resistance level within the entire frequency range b There might be interference in the medical device functioning if the column is being used in the vicinity of equ...

Page 30: ...ective identification plate areas 1 Logo symbol manufacturer production date manufacturer name and address 2 Device symbol 3 General marking CE marking Warning Follow the instructions for safe use Not...

Page 31: ...t 215 kg 39 st 250 kg 6 Maximum load of the extended lying surface frame 27 st 170 kg 7 Note For cleaning washing and disinfecting do not use agents which contain active oxygen or chlorine 8 Infusion...

Page 32: ...st inclination adjustment pedal option 18 Seat inclination adjustment pedal option 19 Warning Follow the instructions for safe use 20 Follow the instructions Note The manufacturer reserves the right t...

Page 33: ...00 3 Famed ywiec Sp z o o 6 Extension cover 1 C220920000003 B00 3 Famed ywiec Sp z o o 7 Lying surface mattress 1 C220920100000 A00 3 Famed ywiec Sp z o o 8 Extension mattress 1 C220920110000 3 Famed...

Reviews: