Instruction Manual for the Delivery Table FREYA 3
Page 7
1.
Proper use and application
1.1.
Application
The delivery table LM-03 is intended for lying and sitting, for patients at all the stages of labour. This
table comes equipped with the permanent and extended lying surfaces. The lying surface height, and
angles of the backrest and seat sections, Trendelenburg and reverse Trendelenburg, are electrically
adjusted.
1.2.
General requirements
The device is intended for indoor use. Required climatic conditions: temperature from +10 to +40°C,
permissible ambient temperature fluctuation within 8 h: below 20°C, relative air humidity: from 30 to
80%, atmospheric pressure: from 700 to 1060 hPa. The device should be used, operated and
maintained acc. to the principles specified in this manual.
NOTE!
Using, operating and servicing of the device in contravention of this instruction
manual is prohibited and may lead to damage, sustained by the user, for which the
manufacturer assumes no liability.
Mounting and installing any equipment that is different from that offered by the
manufacturer is permitted only on the basis of the manufacturer's written consent.
1.3.
User responsibilities
The user – any entity holding an object as an owner, tenant, pledgee, lessee or any entity having a
different right to the object, further – any entity using the device himself or on behalf of whom the
device is used.
The user is obliged to ensure that the device is operated in compliance with its intended purpose and
that it is operated under appropriate conditions, in a manner pursuant to this instruction manual. The
user is obliged to take all necessary measures to ensure safe operation of the device and to reduce all
safety threats to life and health of both user and patient, as well as third parties. Any device operation
is permitted only if by authorised persons, suitably trained and acquainted with this instruction manual.
The user must ensure that all device operators read, understand and observe this instruction manual.
1.4.
Technical data
dimensional length of the delivery table
85 in (2,165 mm) ± 0.8 in (20 mm)
dimensional length of the delivery table in the armchair position
70 in (1,780 mm) ± 0.8 in (20 mm)
max. width (including the rails)
38 in (965 mm) ± 0.2 in (5 mm)
min. height of the lying surface frame
23 in (580 mm) ± 0.4 in (10 mm)
max. height of the lying surface frame
39 in (980 mm) ± 0.4 in (10 mm)
seat elevation angle
30° ± 5°
backrest elevation angle
70° ± 5°
Trendelenburg
20° ± 2°
reverse Trendelenburg
20° ± 2°
delivery table weight
ca. 25 st (160 kg)
permissible load
ca. 40 st (250 kg), that is patient: 34 st (215 kg), and
equipment: 6 st (35 kg)
permissible load of the extended lying surface
27 st (170 kg)
power supply
230 V~ (optionally 100V~, 120V~), 50/60Hz
power consumption
max. 320VA/230V
protection class
II
application part type
B
protection degree
at least IP-X4
life span
10 years
On special request, we can manufacture the device with different technical parameters, without
compromising its safety.