Famed FREYA 3 Instruction Manual Download Page 13

Instruction Manual for the Delivery Table FREYA 3 

Page 13 

3. 

Operation  

 

 

NOTE! 

During  normal  use,  the  table  must  obligatorily  be  connected  to  the  mains. 
Additionally,  the  proper  connection  is  indicated  by  a  lit  indicator  (in  the  option 
where it is included), located in the base cover, above the power cord output from 
the delivery table base. 

3.1. 

Description of the electronic control system 

The  delivery  table  LM-03  comes  equipped  with  an  electronic  control  system,  allowing  the  user  to 
adjust these parameters: 

lying surface height, 

longitudinal tilting angle of the lying surface (Trendelenburg and reverse Trendelenburg), 

backrest section inclination angle 

thigh section inclination angle 

The electrical system control depends on the version: 

central control panel, 

wired controller with the locking functionality 

control panels in the rails 

foot controllers (option) 

 

 

NOTE! 

Control by monostable buttons (require constant activation). 

Each function is activated by pushing the control button and deactivated by releasing it. 
 

3.1.1.  Central control panel 

The central control panel (hung on the high board, fig. 1, item 8c) is intended for the medical staff to 
control  the  table  functions.  It  also  enables  the  staff  to  lock  selected  functions,  so  as  to  prevent  the 
patient from their activation with the wired controller or the ones in the side rails. 

The central control panel enables the control of the following delivery table functions: 

- lying surface height adjustment [6], 
- backrest section inclination angle adjustment [1], 
- thigh section inclination angle adjustment [3], 
- inclination angle adjustment of the backrest and thigh sections (AUTOCONTOUR), [5], 
- lying surface inclination angle adjustment [7], 
- position "cardiac chair" [14], 
- examination position [15], 
- Fowler's position [16], 
- Trendelenburg position [13], 
- zero position [12]. 

To lock the table functions, use the knobs [2, 4, 8, 9].   
The  knob  [2]  locks  the  inclination  angle  adjustment  of  the  backrest  section.  The  knob  [4]  locks  the 
inclination  angle  adjustment  of  the  thigh  section.  The  knob  [8]  locks  the  height  and  tilting  angle 
adjustment  of  the  lying  surface.  The  knob  [9]  locks  all  the  functions.    When  they  are  all  locked,  the 
diode [10] is lit.  
In order to lock the function, twist the proper knob right (its indicator should display the closed padlock 
symbol). In order to unlock the function, twist the proper knob left (its indicator should display the open 
padlock  symbol).  The  buttons  [11]  are  intended  for  the  activation  of  the  functions  despite  their  being 
locked with the knobs. So as to activate the function despite its being locked, push and hold down one 
of the buttons [11] and button assigned to this function.  
 
 

Summary of Contents for FREYA 3

Page 1: ...INSTRUCTION MANUAL Delivery table type LM 03 FREYA 3 Serial no Revision 2 00 Publication date 26 03 2015 FAMED YWIEC Sp z o o Annexes 2 3 4 Index of Spare Parts...

Page 2: ...e essential requirements contained in Annex I to the directive and Act on medical devices The compliance procedure has been performed acc to Annex VII to the directive Manufacturer FAMED YWIEC Sp z o...

Page 3: ...ranty null and void Before any repairs disconnect the device from the mains Safety notes Below is the warning Warning Follow the instructions for safe use This warning can be found on parts and mechan...

Page 4: ...ly exercise caution while adjusting the extended lying surface in relation to the permanent one as the distance between the permanent frame and the extended one is too small if the pressure knob is us...

Page 5: ...r underneath must be free of obstacles During the table movement avoid collisions Exercise caution while removing and mounting the permanent lying surface moulding Application of excessive force may l...

Page 6: ...8 ARMCHAIR POSITION OF THE DELIVERY TABLE CARDIAC CHAIR 18 3 9 QUICK LEVELLING OF THE THIGH AND BACKREST SECTIONS CPR FUNCTION 18 3 10 POSITION 0 18 3 11 EXTENDED LYING SURFACE FRAME POSITION ADJUSTME...

Page 7: ...used The user is obliged to ensure that the device is operated in compliance with its intended purpose and that it is operated under appropriate conditions in a manner pursuant to this instruction man...

Page 8: ...nel with the option of locking the respective electrical functions of the delivery table Control panel in the side rail Wired controller Foot controllers option 9 Backlight lamp from underneath the de...

Page 9: ...related to the structure technology and materials without compromising the technical utility parameters and safety of the device 1 7 Safety The device structure ensures safe operation and use provide...

Page 10: ...nvironment is described in which the medical device Delivery Table LM 03 should be operated The user should observe the guidelines and warnings included in the tables NOTE The use of accessories addit...

Page 11: ...ransport method and package type ensuring safe transport 2 2 Unpacking and first start up NOTE If the device is mounted by the service team of FAMED YWIEC Sp z o o then the user is exempt from perform...

Page 12: ...r the electronic table control system works performing all the functional table movements by use of all the available control elements foot controller and side panels Furthermore check whether the loc...

Page 13: ...elected functions so as to prevent the patient from their activation with the wired controller or the ones in the side rails The central control panel enables the control of the following delivery tab...

Page 14: ...at the same time inclination angle of the backrest and thigh sections AUTOCONTOUR Trendelenburg and reverse Trendelenburg angles The table controller without the central control panel or locking pane...

Page 15: ...with control panels on the outer side see Fig 2a whereas on the inner side with control panels see Fig 2b which enable the activation of the table functions by the patient and medical staff as describ...

Page 16: ...e Trendelenburg when you lower the lying surface to the min height or during the Trendelenburg position marking compliant with pt 8 2 3 3 Backrest section angular position adjustment with the CPR func...

Page 17: ...central panel are marked with the symbol Angular couch position can be changed only by medical personnel The Trendelenburg position is a rescue position NOTE Do not lay any objects on the plastic base...

Page 18: ...e patient s head and push the backrest section down So as to level the thigh section pull the lever under the shank section and push the thigh section down 3 10 Position 0 The position 0 of the delive...

Page 19: ...e between the mounting element of the hand holder and the side rail Failure to observe the above note may at the extreme angles lead to crushing the patient s operator s fingers or to damaging the del...

Page 20: ...r underneath must be free of obstacles During the table movement avoid collisions Do not roll the delivery table across electrical cables 3 15 Use of the side rails Fig 4 The side rails fig 4 are atta...

Page 21: ...ht be damaged Application of excessive force might result in damaging the mouldings 3 17 Infusion rod The infusion rod is embedded in bushings located in the permanent lying surface frame near the hea...

Page 22: ...ually with the agents recommended in annex 2 to this instruction manual NOTE Use foil underlays for spots exposed to preparations applied during various procedures The contact with preparations for di...

Page 23: ...Instruction Manual for the Delivery Table FREYA 3 Page 23 4 Utility parameters The schematic figure below presents the most important utility functions of the delivery table and their marking...

Page 24: ...tion contained herein do not operate it Consult with the manufacturer supplier dealer Using the defective device may lead to damage sustained by the user for which the manufacturer assumes no liabilit...

Page 25: ...rotection NOTE The device contains products which may pose a threat to the environment gel battery type EP 1 2 12 2 pcs Rules of conduct as regards worn products that may pose a threat to the environm...

Page 26: ...should be settled acc to pt 6 6 Troubleshooting So as to ensure safe and reliable functioning within the device lifetime the user shall perform inspections as regards the compliance with standard IEC...

Page 27: ...pply cable disconnected controller defective control system or actuator connect the controller notify an authorised service centre seat section angle adjustment actuator will not work disconnected pow...

Page 28: ...ng features The user is to ensure that the medical device Delivery Table LM 03 is used in this type of environment Resistance IEC 60601 1 2 Test level Resistance level Explanations and guidelines Elec...

Page 29: ...a should be lower than the column resistance level within the entire frequency range b There might be interference in the medical device functioning if the column is being used in the vicinity of equ...

Page 30: ...ective identification plate areas 1 Logo symbol manufacturer production date manufacturer name and address 2 Device symbol 3 General marking CE marking Warning Follow the instructions for safe use Not...

Page 31: ...t 215 kg 39 st 250 kg 6 Maximum load of the extended lying surface frame 27 st 170 kg 7 Note For cleaning washing and disinfecting do not use agents which contain active oxygen or chlorine 8 Infusion...

Page 32: ...st inclination adjustment pedal option 18 Seat inclination adjustment pedal option 19 Warning Follow the instructions for safe use 20 Follow the instructions Note The manufacturer reserves the right t...

Page 33: ...00 3 Famed ywiec Sp z o o 6 Extension cover 1 C220920000003 B00 3 Famed ywiec Sp z o o 7 Lying surface mattress 1 C220920100000 A00 3 Famed ywiec Sp z o o 8 Extension mattress 1 C220920110000 3 Famed...

Reviews: