Instruction Manual for the Delivery Table FREYA 3
Page 17
3.4.
Seat section angular position adjustment
The seat angular position is adjusted by an electric actuator (fig. 1, item 8b). The seat section stops
automatically on reaching the extreme angles.
Control method:
♦
central control panel – (fig. 1, item 8c),
♦
control panel in the side rail (fig. 1, item 8d)
♦
wired controller (fig.1, item 8e)
♦
foot control pedal (option) (fig. 1, item 8f)
The buttons for the seat angular position adjustment and knob for
locking this function on the central panel are marked with the symbol:
NOTE!
Please, do acquaint yourself with the safety warnings, contained in the
introduction to this manual.
3.5.
Lying surface angular position adjustment
The lying surface angular position (Trendelenburg and reverse Trendelenburg) of the delivery table is
adjusted by an electric actuator. The lying surface stops automatically on reaching the extreme angles.
Control method:
♦
central control panel – (fig. 1, item 8c),
♦
control panel in the side rail (fig. 1, item 8d)
♦
wired controller (fig.1, item 8e)
♦
foot control pedal (option) (fig. 1, item 8f)
The buttons for the lying surface angular position adjustment and knob for
locking this function on the central panel are marked with the symbol:
Angular couch position can be changed only by medical personnel. The
Trendelenburg position is a rescue position.
NOTE!
Do not lay any objects on the plastic base cover, since it might damage: the base
cover, lying surface elements, and angular position adjustment function of the lying
surface – Trendelenburg, when you lower the lying surface to the min. height or
during the Trendelenburg position, marking compliant with pt. 8.2.
3.6.
Autocontour function
The autocontour function enables the angular position adjustment for both backrest and thigh
sections, simultaneously. The actuator's movement is activated only when the button is pressed. After
its release, the movement stops. The sections stop automatically on reaching the extreme angles.
The buttons for the control of the autocontour function on the control system
elements are marked with the symbol:
NOTE!
Exercise caution while activating this function. The movement of the thigh and
backrest sections (caused by actuators) may result in clipping your fingers or
hand. Do not press your hand against the table elements near the moving
elements of the structure.
If there is a risk that the patient's condition might lead to trapping the patient,
then the lying surface sections should remain levelled, in the absence of medical
supervision (except when the patient's condition does not allow it).