Instruction Manual for the Delivery Table FREYA 3
Page 4
press your hand against the table elements near the moving elements of the
structure.
•
If the patient lies on the table without being supervised by the medical staff, then
the table should remain in the lowest position.
•
If the patient is not attended to by the medical staff, lock the table functions
activation of which may pose a threat to the patient safety.
•
If there is a risk that the patient's condition (e.g. disorientation due to medical
reasons or her clinical state) might lead to trapping the patient, then the lying
surface ought to remain flat, in the absence of supervision (unless otherwise
advised by the medical personnel, or under special/particular circumstances).
•
Avoid safety risks which may result from the improper use of power cables/cords,
e.g. tensioning the cables/cords, their squeezing in between the moving parts,
running over them.
•
Do not roll the delivery table across electrical cables!
The patient's weight and additional equipment's, mounted at the table, may not
exceed 40 st (250 kg).
Maximum permissible load of the extended lying surface: up to 27 st (170 kg).
Obligatorily exercise caution while adjusting the extended lying surface in relation
to the permanent one, as the distance between the permanent frame and the
extended one is too small (if the pressure knob is used). In extreme cases, it may
lead to crushing the patient's or user's fingers.
Maximum permissible working load of the foot support: up to 500 N.
If the device is to be used as a standard bed, mount the footboard.
Maximum permissible working load of the calf support, mounted in the fastening
holder, on the fastening strip: up to 5 st (30 kg).
Maximum permissible working load of the hand holder, mounted in the multi-
position holder, on the fastening strip: up to 500 N.
Maximum permissible working load of all the infusion rod hooks: up to 0.5 st
(3 kg).
The delivery table is intended for patients whose height does not exceed 79 in
(200 cm).
Start-up, operation and use notes
During normal use, the delivery table must obligatorily be connected to the mains.
The bed, during use, should not be in the place which obstructs its disconnection
from the mains!
If you use the table near high-frequency medical apparatus and defibrillators,
carefully observe recommendations contained in instruction manuals of this
equipment. Improper use may result in dangerous accidents. If the patient touches
the metal table parts or equipment, she runs the risk of serious burning.