
Instruction Manual for the Delivery Table FREYA 3
Page 26
Technical inspection scope
Frequency
-
functionality check
-
general technical state check
every 12 months
So as to ensure safe and reliable functioning within the device lifetime, the user shall perform technical
state inspections every six months acc. to the table below.
Technical state inspection range
Undercarriage
check of the bolted joints and protective elements on the rotary pivots
check whether the central locking device and chassis work
lubricate the rotary knots with the grease OBEEN UF3 by OPTIMOL (distributor:
CASTROL)
Lying surface
check of the electrical actuators' operation
check of the CPR function's operation
check of the bolted joints and protective elements on the rotary pivots
check of the clamp and guides of the extended lying surface
lubricate the rotary knots with the grease OBEEN UF3 by OPTIMOL (distributor:
CASTROL)
Checks should be performed through visual inspection. Non-conformities should be settled acc. to pt.
6.6 Troubleshooting.
So as to ensure safe and reliable functioning within the device lifetime, the user shall perform
inspections as regards the compliance with standard IEC 62353, acc. to the table below.
The tests are carried out in the following situations:
- after repair
- periodically (every 12 months)
Inspection range
Visual
inspection
•
Check of all the fuses available from the outside, their compliance with the
documentation (rated current, characteristics),
•
Check for the presence and readability of all the markings and labels,
•
Check for the presence of all the mechanical parts,
•
Check for damage and cleanliness,
•
Check of the accessories,
•
Check of the documentation of the device and earlier inspections.
Measurements
•
Measurement of the protective grounding resistance,
•
Measurement of leakage currents,
•
Measurement of the patient's leakage current,
•
Measurement of the insulation resistance.
Functional test
•
The functional test should be carried out in the presence of a person who is
knowledgeable and trained in terms of the particular device's application,
•
The functional test and results of the respective trials and measurements
should be documented.