102
Connect 30.0VDC and a voltmeter to Probe-HI + and Probe-LO -. When the power supply and
meter are connected, press TEST to activate the calibration. Read the measurement from your
standard and enter it using the numeric keypad. You may now store the new calibration
number by pressing the ENTER key or escape by pressing the EXIT key or the RESET button.
9.
LLT
×
1 DC range
:
Calibration of LLT MD Volts DC x 1
Connect 8.0VDC and a voltmeter to Probe-HI + and Probe-LO -. When the power supply and
meter are connected, press TEST to activate the calibration. Read the measurement from your
standard and enter it using the numeric keypad. You may now store the new calibration
number by pressing the ENTER key or escape by pressing the EXIT key or the RESET button.
10.
LLT
×
4 DC range
:
Calibration of LLT MD Volts DC x 4