34
3.4 OPT.753 HV & GB Link Module
1. PROBE HI / PROBE LO
Probe HI terminal is an input to one side of the MD(measuring device) and will be enabled
during a Touch Current test when ever Probe-HI has been selected at setup. This terminal is
provided for performing Enclosure leakage or Applied Part leakage tests. When run tests and
Touch Current tests are not being performed, this terminal is isolated from the run test and
Touch Current test circuits.
Probe LO terminal is an input to one side of the MD(measuring device) and will be enabled
during a Touch Current test when ever Probe-LO has been selected at setup. This terminal is
provided for performing Applied Part leakage tests and is always used in conjunction with
the Probe-HI terminal. When run tests and Touch Current tests are not being performed, this
terminal is isolated from the run test and Touch Current test circuits.
2. GND TERMINAL
This terminal is where the Ground or Earth terminal from the adapter box is connected.
3. CASE TERMINAL
This terminal is connected to the DUT case or dead metal and provides the return for the
Ground Bond, Dielectric Withstand, and Insulation Resistance tests. During a Run test or
Touch Current test, this terminal is isolated from the test circuits.
4. CURRENT OUTPUT JACK
Connector used to attach the high current output lead, adapter box high current lead or test
fixture high current lead to the instrument. This connection provides the output current for
the ground bond and continuity.
1 2 3 4
5 6