99
Set DUT I/P 120-277VAC. Connect a 40Amp load at DUT output in series with current meter.
When the load is connected, press TEST to activate the calibration. Read the measurement
from your standard and enter it using the numeric keypad. You may now store the new
calibration number by pressing the ENTER key or escape by pressing the EXIT key or the
RESET button.
3. RUN TEST 4.000A
:
Calibration of Run Test Current 4.000A
Set DUT I/P 120-277VAC. Connect a 4Amp load at DUT output in series with current meter.
When the load is connected, press TEST to activate the calibration. Read the measurement
from your standard and enter it using the numeric keypad. You may now store the new
calibration number by pressing the ENTER key or escape by pressing the EXIT key or the
RESET button.
4.RUN TEST 10kW
:
Calibration of Run Test Power 10kW