56
Power-HI Power-LO
Enter the new power via the numeric keypad and then press the ENTER key to accept the new
parameter or press the EXIT key to escape from the edit.
4.3.2.6 Power Factor (PF-HI and PF-LO)
Please use the up and down arrow keys or the ENTER key and scroll the highlighted area to
the
PF-HI
or
PF -LO
parameter. The display will appear as follows:
PF-HI PF-LO
Enter the new PF via the numeric keypad and then press the ENTER key to accept the new
parameter or press the EXIT key to escape from the edit.
Power Factor = W/VA where W =Watts (Real Power) and VA =Volts x Amps (apparent power).
It is important to note that the closer the power factor is to "1" the more resistive the DUT is.
The closer the power factor is to 0 the more reactive (inductive or capacitive) the DUT is.
4.3.2.7 PLC Control (Option for Opt. 752)
Please use the up and down arrow keys or the ENTER key and scroll the highlighted area to
the
PLC Control
parameter.
Please press
PLC Control
softkey to select the power mode. If option AC power source to be
the DUT input power, there are 7 memories M1 to M7 for selected. If option 1931 series, there
are three voltage rate 1.0, 1.06, and 1.1 times for selected. If option 1931S series, there are six
voltage rate 0.8, 0.85, 0.9, 1.0, 1.06, and 1.1 times for selected.
4.3.2.8 Continuous
The softkey can set the output power status for step link. When Continuous is on, and both two
link step are at the same line status, the power will continue output during two steps link. If
Continuous is on but two link step are not at the same line status, the power will stop output