113
Connect 2.0VD
C in series with a current meter to Probe-HI + and Probe-LO -. When the
power supply and meter are connected, press TEST to activate the calibration. Read the
measurement from your standard and enter it using the numeric keypad. You may now store
the new calibration number by pressing the ENTER key or escape by pressing the EXIT key
or the RESET button.
31.LLT IEC60990
FIG 4
:
Calibration of Touch Current Test MD: IEC60990 FIG 4
measuring device
Connect 2.0
VDC in series with a current meter to Probe-HI + and Probe-LO -. When the
power supply and meter are connected, press TEST to activate the calibration. Read the
measurement from your standard and enter it using the numeric keypad. You may now store
the new calibration number by pressing the ENTER key or escape by pressing the EXIT key
or the RESET button.
32.LLT IEC60990
FIG5
:
Calibration of Touch Current Test MD: IEC60990 FIG 5
measuring device