32
Rear Panel Controls (Continuing)
8. INTERFACE
Standard connector for interconnection to the RS-232 Bus interface. Optional IEEE 488
interface or Opt.751 Multi-function Interface Card may be substituted for the RS-232.
Multi-function interface card including USB port A type, combination of RS-232,
RS-485 port, and Bar Code Input with internal memory capability.
9. INPUT POWER RECEPTACLE
Standard IEC 320 connector for connection to a standard NEMA style line power (mains)
cord.
10. INPUT POWER SWITCH
Line voltage selection is set by the position of the switch. In the up position it is set for 115
volt operation, in the down position it is set for 230 volt operation.
11. REMOTE SIGNAL INPUT
9-Pin D subminiature male connector for remote control of test, reset, and interlock functions,
as well as remote program file selection.
12. REMOTE SIGNAL OUTPUT
9-Pin D sub-miniature female connector for monitoring PASS, FAIL, and PROCESSING
output relay signals.
13.
CALIBRATION BUTTON
To put the instrument into the calibration mode, push this button and turn on the power switch
simultaneously.
10
17
16
15
14
12
11
1 2
3
6 7 8
9
4
5
13