Installation
GMS
plus
5.
Setting the span using a
comparative measurement
Follow the same procedure as factory setting; instead of executing
a test, the comparative measurement value is set.
Function test with chlorine gas
generator
Proceed as follows:
1
Start the chlorine gas generator (Part No. W3T159905).
Wait 30 minutes for the generator to run in.
Please note
Read operating instructions for the gas generator!
2
Place the 1 ppm chlorine disc into the air inlet of the gas
generator.
3
Place the sensor into the chlorine gas generator.
Once the display on the gas monitoring system GMS
plus
has
stabilized, the displayed value should be 1 ppm ± 50 %.
4
Switch off the chlorine gas generator.
5
Remove the sensor.
Summary of Contents for WALLACE & TIERNAN GMS PLUS
Page 1: ...WALLACE TIERNAN GAS MONITORING SYSTEM GMS PLUS INSTRUCTION MANUAL...
Page 2: ...2 WT 050 455 000 DE IM 0319 GMS plus Please note Original instruction manual...
Page 20: ...20 WT 050 455 000 DE IM 0319 Description GMS plus 3...
Page 38: ...38 WT 050 455 000 DE IM 0319 Functions GMS plus 4...
Page 76: ...76 WT 050 455 000 DE IM 0319 Operation GMS plus 6 General overview...
Page 94: ...94 WT 050 455 000 DE IM 0319 Operation GMS plus 6...
Page 103: ...GMS plus Wiring diagrams 9 103 9 Wiring diagrams...
Page 104: ...104 WT 050 455 000 DE IM 0319 Wiring diagrams GMS plus 9...
Page 105: ...GMS plus Wiring diagrams 9 105...
Page 106: ...106 WT 050 455 000 DE IM 0319 Wiring diagrams GMS plus 9...
Page 107: ...GMS plus Wiring diagrams 9 107...
Page 108: ...108 WT 050 455 000 DE IM 0319 Wiring diagrams GMS plus 9...
Page 109: ...GMS plus Declaration of conformity 10 109 10 Declaration of conformity...
Page 110: ...110 WT 050 455 000 DE IM 0319 Declaration of conformity GMS plus 10...
Page 114: ...114 WT 050 455 000 DE IM 0319 Index GMS plus 11...