
56
WT.050.455.000.DE.IM.0319
Installation
GMS
plus
5.
Setting the span using a
comparative measurement
Follow the same procedure as factory setting; instead of executing
a test, the comparative measurement value is set.
Function test with ozone gas
generator
Proceed as follows:
1
Start the ozone gas generator (Part No. W3T171363).
Wait 5 minutes for the generator to run in.
Please note
Read operating instructions for the gas generator!
2
Place the sensor into the ozone gas generator. Once the
display on the gas monitoring system GMS
plus
has stabilized,
the displayed value should be 0,4 ppm ± 25 %.
3
Switch off the ozone gas generator.
4
Remove the sensor.
Summary of Contents for WALLACE & TIERNAN GMS PLUS
Page 1: ...WALLACE TIERNAN GAS MONITORING SYSTEM GMS PLUS INSTRUCTION MANUAL...
Page 2: ...2 WT 050 455 000 DE IM 0319 GMS plus Please note Original instruction manual...
Page 20: ...20 WT 050 455 000 DE IM 0319 Description GMS plus 3...
Page 38: ...38 WT 050 455 000 DE IM 0319 Functions GMS plus 4...
Page 76: ...76 WT 050 455 000 DE IM 0319 Operation GMS plus 6 General overview...
Page 94: ...94 WT 050 455 000 DE IM 0319 Operation GMS plus 6...
Page 103: ...GMS plus Wiring diagrams 9 103 9 Wiring diagrams...
Page 104: ...104 WT 050 455 000 DE IM 0319 Wiring diagrams GMS plus 9...
Page 105: ...GMS plus Wiring diagrams 9 105...
Page 106: ...106 WT 050 455 000 DE IM 0319 Wiring diagrams GMS plus 9...
Page 107: ...GMS plus Wiring diagrams 9 107...
Page 108: ...108 WT 050 455 000 DE IM 0319 Wiring diagrams GMS plus 9...
Page 109: ...GMS plus Declaration of conformity 10 109 10 Declaration of conformity...
Page 110: ...110 WT 050 455 000 DE IM 0319 Declaration of conformity GMS plus 10...
Page 114: ...114 WT 050 455 000 DE IM 0319 Index GMS plus 11...