GMS
plus
Installation
5.
45
5.2.3
Electrical installation
See 9. “Wiring diagrams” on page 103.
Warning!
Risk involving electric current.
Only authorized and qualified electricians are permitted to install
the unit and open the housing. The unit may only be taken into
operation when the housing is closed, and must be connected to
protective earth. Modifications to the unit that go beyond those
described in this manual are not permissible.
Comply with local regulations when installing the unit (e. g. GUV).
Warning!
The unit is not equipped with a mains switch and is in operation as
soon as the supply voltage is applied. An external switch or circuit
breaker is therefore necessary.
Provide a mains fuse locally (6 A). The conductor cross section of
the mains cable must be at least 0.75 mm (AWG 18).
When connecting system components (e.g. devices, motors,
pumps) as well as when entering operating data, the system
components must be switched off in order to prevent uncontrolled
activation or any incorrect function.
Attention!
To ensure safe and correct commissioning, knowledge of the
operation, connected electrical load, measurement signals, cable
assignment and fuse protection of the connected devices and
machines and the relevant safety regulations is required.
Commissioning of the unit may therefore only be performed by
qualified and authorized electricians.
Incorrectly connected devices can be damaged, possibly
irreparably, or cause faults in other equipment when they are
switched on or in operation. Ensure that the measuring and control
cables are not mixed up or are able to make contact with one
another. Never connect or disconnect any cables to which voltage
is applied!
Summary of Contents for WALLACE & TIERNAN GMS PLUS
Page 1: ...WALLACE TIERNAN GAS MONITORING SYSTEM GMS PLUS INSTRUCTION MANUAL...
Page 2: ...2 WT 050 455 000 DE IM 0319 GMS plus Please note Original instruction manual...
Page 20: ...20 WT 050 455 000 DE IM 0319 Description GMS plus 3...
Page 38: ...38 WT 050 455 000 DE IM 0319 Functions GMS plus 4...
Page 76: ...76 WT 050 455 000 DE IM 0319 Operation GMS plus 6 General overview...
Page 94: ...94 WT 050 455 000 DE IM 0319 Operation GMS plus 6...
Page 103: ...GMS plus Wiring diagrams 9 103 9 Wiring diagrams...
Page 104: ...104 WT 050 455 000 DE IM 0319 Wiring diagrams GMS plus 9...
Page 105: ...GMS plus Wiring diagrams 9 105...
Page 106: ...106 WT 050 455 000 DE IM 0319 Wiring diagrams GMS plus 9...
Page 107: ...GMS plus Wiring diagrams 9 107...
Page 108: ...108 WT 050 455 000 DE IM 0319 Wiring diagrams GMS plus 9...
Page 109: ...GMS plus Declaration of conformity 10 109 10 Declaration of conformity...
Page 110: ...110 WT 050 455 000 DE IM 0319 Declaration of conformity GMS plus 10...
Page 114: ...114 WT 050 455 000 DE IM 0319 Index GMS plus 11...