88
WT.050.455.000.DE.IM.0319
Operation
GMS
plus
6.
Display
Value range
(factory setting, bold)
Description
DIAGNOSIS
Error diagnostics
Input I
for mA sensors in mA
for measuring range start => 4mA
for measuring range end => 20mA
for sensor error < 4 mA
Chlorine:
for 0ppm => approx. 1µA
for 5ppm => 16.5µA
for 20ppm => 58.7µA
for sensor error <0.1µA
Chlorine Dioxide:
for 0ppm => approx. 1µA
for 5ppm => 9µA
for 20ppm => 30.9µA
for sensor error <0.1µA
Ozone:
for 0ppm => approx. 0.3µA
for 5ppm => 1.6µA
for sensor error <0.1µA
The current sensor signal is displayed
for Chloratect sensors in µA
for mA sensors in mA
Input II
See Input I
See Input I
Input Temp I
13°C
Temperature on temperature sensor I
(uncalibrated)
Input Temp II
25°C
Temperature on temperature sensor II
(uncalibrated)
REL: 1 2 3 4 5 6 7 8
Rel 1 Alarm 1
Rel 2 Alarm 2 I
Rel 3 Alarm 1 II
Rel 4 Alarm 2 II
Rel 5 Common Alarm 1
Rel 6 Common Alarm 2
Rel 7 Common Alarm 3
Rel 8 Sensoralarm
0 0 0 0 0 0 0 0
Current signal condition at relay outputs 1 –
8
0 = relay off
1 = relay on
Summary of Contents for WALLACE & TIERNAN GMS PLUS
Page 1: ...WALLACE TIERNAN GAS MONITORING SYSTEM GMS PLUS INSTRUCTION MANUAL...
Page 2: ...2 WT 050 455 000 DE IM 0319 GMS plus Please note Original instruction manual...
Page 20: ...20 WT 050 455 000 DE IM 0319 Description GMS plus 3...
Page 38: ...38 WT 050 455 000 DE IM 0319 Functions GMS plus 4...
Page 76: ...76 WT 050 455 000 DE IM 0319 Operation GMS plus 6 General overview...
Page 94: ...94 WT 050 455 000 DE IM 0319 Operation GMS plus 6...
Page 103: ...GMS plus Wiring diagrams 9 103 9 Wiring diagrams...
Page 104: ...104 WT 050 455 000 DE IM 0319 Wiring diagrams GMS plus 9...
Page 105: ...GMS plus Wiring diagrams 9 105...
Page 106: ...106 WT 050 455 000 DE IM 0319 Wiring diagrams GMS plus 9...
Page 107: ...GMS plus Wiring diagrams 9 107...
Page 108: ...108 WT 050 455 000 DE IM 0319 Wiring diagrams GMS plus 9...
Page 109: ...GMS plus Declaration of conformity 10 109 10 Declaration of conformity...
Page 110: ...110 WT 050 455 000 DE IM 0319 Declaration of conformity GMS plus 10...
Page 114: ...114 WT 050 455 000 DE IM 0319 Index GMS plus 11...