GMS
plus
Installation
5.
57
5.5
Commissioning the mA sensor
Remove the protection cap
Please note
The sensor opening protection cap must be removed before
commissioning the sensor!
Proceed as follows:
1
Remove the sensor opening protection cap.
Installing the mA sensor
2
Secure the sensor cable (2 m) or sensor cable (10 m) to
terminal 19 (red) and 20 (blue).
Channel II => terminal 23 (red) and terminal 24 (blue).
3
Connect the BNC plug into the mA sensor.
Please note
During the run-in time the sensor adapts to the environmental
conditions.
4
In the menu"SENSOR I", parameter "Sensor input I" , set the
"mA Sensor".
5
In the menu "Sensor type I", set sensor types"Cl2", "ClO2",
"O3", etc. and/or select the variably definable sensor type"----"
.
You may define four characters for the variable sensor type.
6
In the menu "Sensor unit I", select the desired unit or variably
definably unit "----".
You may define four characters for the variable unit.
7
Set the measuring range-initial value (see sensor data sheet).
A
B
C
A
Sensor opening
B
Protection cap
C
Base
A
A BNC plug
Summary of Contents for WALLACE & TIERNAN GMS PLUS
Page 1: ...WALLACE TIERNAN GAS MONITORING SYSTEM GMS PLUS INSTRUCTION MANUAL...
Page 2: ...2 WT 050 455 000 DE IM 0319 GMS plus Please note Original instruction manual...
Page 20: ...20 WT 050 455 000 DE IM 0319 Description GMS plus 3...
Page 38: ...38 WT 050 455 000 DE IM 0319 Functions GMS plus 4...
Page 76: ...76 WT 050 455 000 DE IM 0319 Operation GMS plus 6 General overview...
Page 94: ...94 WT 050 455 000 DE IM 0319 Operation GMS plus 6...
Page 103: ...GMS plus Wiring diagrams 9 103 9 Wiring diagrams...
Page 104: ...104 WT 050 455 000 DE IM 0319 Wiring diagrams GMS plus 9...
Page 105: ...GMS plus Wiring diagrams 9 105...
Page 106: ...106 WT 050 455 000 DE IM 0319 Wiring diagrams GMS plus 9...
Page 107: ...GMS plus Wiring diagrams 9 107...
Page 108: ...108 WT 050 455 000 DE IM 0319 Wiring diagrams GMS plus 9...
Page 109: ...GMS plus Declaration of conformity 10 109 10 Declaration of conformity...
Page 110: ...110 WT 050 455 000 DE IM 0319 Declaration of conformity GMS plus 10...
Page 114: ...114 WT 050 455 000 DE IM 0319 Index GMS plus 11...