74
EMControl Command Set
ets-lindgren.com
UV
Description:
Changes the upper limit of the device for vertical polarity. The
specified value must be greater than the lower limit for vertical
polarization and between
–999 and 999.
Devices:
Towers
Syntax:
UV [+/
–] XXX
+
: Optional for positive values
–
: Necessary for negative values
XXX
: Decimal value for the upper limit to be set in
centimeters. Leading zeroes are optional.
Response:
OK
, or an error (error number)
Example:
Write
sdUV 355\n
: Set vertical polarity upper limit to 355 cm
See Also:
UV?
UV?
Description:
Queries the upper limit of the device for vertical polarity. The
value returned is in XXX format depending on the current
numeric mode. Negative values are preceded by a
–
(minus
sign).
Devices:
Towers
Syntax:
UV?
Response:
[
–]XXX
: Value of the upper limit for the vertical polarization in
centimeters
Example:
Write
sdUV?\n
: Query vertical upper limit
Read response
Response is
350\n
: Vertical upper limit is 350 cm
See Also:
UV
Summary of Contents for ETS-Lindgren EMControl 7006-001
Page 1: ...Model 7006 001 EMControl Positioner Controller Plug In Card User Manual...
Page 8: ...viii ets lindgren com This page intentionally left blank...
Page 12: ...12 Introduction ets lindgren com This page intentionally left blank...
Page 16: ...16 Specifications ets lindgren com This page intentionally left blank...
Page 18: ...18 EMControl Plug In Card Installation ets lindgren com This page intentionally left blank...
Page 88: ...88 EMControl Command Set ets lindgren com This page intentionally left blank...
Page 90: ...90 Warranty ets lindgren com This page intentionally left blank...