Operation
31
Press
Device Name
to enter a custom name
for the device.
Press
Device Type
to select the type of
device: linear (tower) or rotational
(turntable). For rotational devices, a
Continuous
or
non-Continuous
mode
sub-type can be selected; for linear devices
a
Tower
or
Bore Sight Tower
sub-type can
be selected
.
In the continuous mode of operation a turntable is allowed
unlimited movement. The counter readout is from 0 to
359.9 and the software limits are ignored. In the
non-continuous mode, the turntable is restricted between
upper and lower software limits.
Note:
All devices with an ETS-Lindgren motor base incorporate a
counter or encoder. The number of counts per meter must be set to
ensure accurate movement. Normally this number is factory-set and
should not be changed.
In jog mode, users can select whether limits are
respected. If
Jog Limits
is set to
Enabled
, movement is
stopped when software limits are reached. If set to
Disabled
, movement can continue beyond software limits.
When
Jog Quick Stop
is disabled, stopping when in
jog mode uses the device acceleration parameter to stop
slowly. If
Jog Quick Stop
is enabled, releasing the button
will result in an immediate stop.
Summary of Contents for ETS-Lindgren EMControl 7006-001
Page 1: ...Model 7006 001 EMControl Positioner Controller Plug In Card User Manual...
Page 8: ...viii ets lindgren com This page intentionally left blank...
Page 12: ...12 Introduction ets lindgren com This page intentionally left blank...
Page 16: ...16 Specifications ets lindgren com This page intentionally left blank...
Page 18: ...18 EMControl Plug In Card Installation ets lindgren com This page intentionally left blank...
Page 88: ...88 EMControl Command Set ets lindgren com This page intentionally left blank...
Page 90: ...90 Warranty ets lindgren com This page intentionally left blank...