68
EMControl Command Set
ets-lindgren.com
TT
Description:
Selects turntable mode of operation. Allows the selection of
normal, two speed, and air flotation turntables.
In the air flotation mode of operation, the UP/DOWN is controlled
automatically by the EMControl, assuring that the table is floating
before a movement is made.
This command is intended to allow automated configuration of
the EMControl and should not be used by most programs.
Devices:
Turntables, towers
Syntax:
TT <NRM/AIR/TWO> <CONT/NONCONT>
NRM
: Normal turntable
AIR
: Air flotation turntable
TWO
: Two-speed turntable
CONT
: Continuous rotation turntable
NONCONT
: Non-continuous rotation turntable
Response:
OK
, or an error (error number)
Example:
Write
sdTT NRM CONT\n
: Change device from a tower to a
normal, continuous turntable
See Also:
TWR
,