ets-lindgren.com
EMControl Command Set
39
CL
Description:
Changes the counterclockwise limit of the device. The specified
value must be less than the clockwise limit and between
–999 and 999.
Devices:
Turntables
Syntax:
CL [+/
–] XXX
+
: Optional for positive values
–
: Necessary for negative values
XXX
: Decimal value for the counterclockwise limit to be set in
degrees. Leading zeroes are optional.
Response:
OK
, or an error (error number)
Example:
Write
sdCL
–100\n
: Change CCW limit of the turntable to
–100 degrees
See Also:
CL?
CL?
Description:
Queries the turntable counterclockwise limit. The value returned
in XXX format. Negative values are preceded by a
–
(minus sign).
Devices:
Turntables
Syntax:
CL?
Response:
[-]XXX
: Value of the counterclockwise limit setting in degrees
Example:
Write
sdCL?\n
: Query turntable counterclockwise limit
Read response
Response is
200\n
: Counterclockwise limit is 200 degrees
See Also:
CL
Summary of Contents for ETS-Lindgren EMControl 7006-001
Page 1: ...Model 7006 001 EMControl Positioner Controller Plug In Card User Manual...
Page 8: ...viii ets lindgren com This page intentionally left blank...
Page 12: ...12 Introduction ets lindgren com This page intentionally left blank...
Page 16: ...16 Specifications ets lindgren com This page intentionally left blank...
Page 18: ...18 EMControl Plug In Card Installation ets lindgren com This page intentionally left blank...
Page 88: ...88 EMControl Command Set ets lindgren com This page intentionally left blank...
Page 90: ...90 Warranty ets lindgren com This page intentionally left blank...