EMControl Command Set
ets-lindgren.com
CY
Description:
Changes the cycle count for the device. The value must be
between 0.0 and 999.5. A value of
0
or
000.0
represents an
infinite scan count.
A cycle represents a full scan from limit to limit and back again
starting at the closest limit. The addition of the 0.5 cycle steps
allows the controller to scan between the limits an integral
number of times instead of always performing an even number of
scans from limit to limit.
Devices:
Towers, turntables
Syntax:
CY XXX.X
XXX.X
: Decimal value between 0.0 and 999.5 in steps of 0.5.
Leading zeroes are optional.
Response:
OK
, or an error (error number)
Example:
Write
sdCY 000\n
: Set cycle count of tower to infinite
Write
sdCY 2.5\n
: Set the turntable to scan between its limits
2.5 times
See Also:
CY?
CY?
Description:
Queries the cycle count for the device. A four-digit value is
returned between the value of 000.0 and 999.5. A value of 000.0
represents an infinite scan count.
Devices:
Towers, turntables
Syntax:
CY?
Response:
XXX.X
: Value of the scan cycle setting
Example:
Write
sdCY?\n
: Query turntable cycle count
Read response
Response is
5\n
: Cycle count is 5
See Also:
CY
Summary of Contents for ETS-Lindgren EMControl 7006-001
Page 1: ...Model 7006 001 EMControl Positioner Controller Plug In Card User Manual...
Page 8: ...viii ets lindgren com This page intentionally left blank...
Page 12: ...12 Introduction ets lindgren com This page intentionally left blank...
Page 16: ...16 Specifications ets lindgren com This page intentionally left blank...
Page 18: ...18 EMControl Plug In Card Installation ets lindgren com This page intentionally left blank...
Page 88: ...88 EMControl Command Set ets lindgren com This page intentionally left blank...
Page 90: ...90 Warranty ets lindgren com This page intentionally left blank...