14 /
48
Chapter 3
W a r n i n g
The O-scan magnet is a permanent magnet and its static magnetic
field cannot be de-activated in an emergency.
Never use resuscitation devices such as defibrillators, oxygen tanks,
etc., within the controlled access area.
Remove the patient from the influence of the magnet, following the
procedure described in the chapter “Positioning Protocols” in this
document.
W a r n i n g
During an examination, if an anomalous shutdown of the OPI -
Operator Interface program occurs, the user must turn off the external
wall-mounted power switch to stop the radio-frequency impulse and
generation of the magnetic field gradient.