350004900 Rev. 07
7
/ 20
•
•
•
•
•
•
T
echnical Description
System electronics
The electronic components are housed in the electronic box, as shown in
the figure below:
fig. 5.3 - Details of electronics
Power supply
It comprises:
1
TRI, an insulation transformer to limit leakage current to earth
according to standard EN 60601-1 and subsequent amendments,
referred to permanently installed medical systems
2
DISTR unit, a distribution box for power of the various modules, fitted
with a network filter
PC
It comprises:
1
a commercial PC, with PCI bus interface
2
SDSP card, which performs the following functions:
•
control of ACM card
•
acquisition and processing of the MR signal and monitoring of data
from the ACM card