Chapter 4
Electromagnetic requirements in the installation environment
Electromagnetic compatibility (EMC)
The O-scan system requires special precautions regarding
electromagnetic compatibility and installation and set-up must be
according to the EMC instructions provided in this section.
The configuration of O-scan required to ensure compliance with the
requirements of electromagnetic compatibility, envisages that the cables
required for connection of the system to the mains, as well as connection
of the system modem to the telephone line must have the properties
described below.
•
LAN connection: Ethernet cable, length less than or equal to 5 m,
with load impedance of 100 Ohm.
N o t e
The system is intended to provide basic safety and essential
performance producing images free of artefacts or distortions that a
medical expert, trained in the use of MR equipment, is not able to
distinguish from the internal structure of the anatomical area under
examination or, as an alternative, not producing images temporarily.
W a r n i n g
Use of a shielded booth and/or connection cables other than as
specified above may cause an increase in emissions or a decrease in
immunity of the O-scan system.
Summary of Contents for O-scan eXP
Page 1: ......
Page 2: ......
Page 3: ...O scan User Manual 350004900 Rev 07 Esaote S p A Genoa...
Page 9: ...350004900 Rev 07 7 10...
Page 10: ...8 10...
Page 12: ...10 10...
Page 20: ...8 8...
Page 32: ...8 8 Chapter 2...
Page 51: ...350004900 Rev 07 19 48 Instructions for Safe Use fig 3 1 Top view...
Page 52: ...20 48 Chapter 3 fig 3 2 Side view...
Page 80: ...48 48 Chapter 3...
Page 102: ...22 22 Chapter 4...
Page 105: ...350004900 Rev 07 3 20 Technical Description fig 5 1 Equipment components...
Page 122: ...20 20 Chapter 5...
Page 142: ...20 20 Chapter 6...
Page 161: ...350004900 Rev 07 5 10 Soft cushion Set fig 9 6 Use of lumbar support cushion...
Page 184: ...10 10 Chapter 11...
Page 232: ...4 4 Chapter 17...
Page 236: ...4 4 Chapter 18...