14 /
22
Chapter 4
Environmental Requirements
Temperature and humidity requirements
To function within its specifications, O-scan must be operated in an
ambient temperature range of 20 °C to 26 °C (68 °F to 78.8 °F), with
fluctuation no greater than ±3 °C (37.4 °F) per hour. The ambient relative
humidity must be in the range of 45% to 80%.
The installation area must be air conditioned to assure the stability of the
temperature and humidity specifications.
Other requirements
A t t e n t i o n
Never install O-scan in locations exposed to direct sunlight, air
draughts and temperature variations caused by other sources.
A t t e n t i o n
Do not install O-scan in locations where:
•it is exposed to steam, dripping water, or a saline atmosphere
• it is exposed to excessively dusty environments
• it is subject to severe vibrations or shocks
• is it subject to inclines of more than 10 degrees
If O-scan is not used for prolonged intervals, cover with a waterproof
sheet.
Summary of Contents for O-scan eXP
Page 1: ......
Page 2: ......
Page 3: ...O scan User Manual 350004900 Rev 07 Esaote S p A Genoa...
Page 9: ...350004900 Rev 07 7 10...
Page 10: ...8 10...
Page 12: ...10 10...
Page 20: ...8 8...
Page 32: ...8 8 Chapter 2...
Page 51: ...350004900 Rev 07 19 48 Instructions for Safe Use fig 3 1 Top view...
Page 52: ...20 48 Chapter 3 fig 3 2 Side view...
Page 80: ...48 48 Chapter 3...
Page 102: ...22 22 Chapter 4...
Page 105: ...350004900 Rev 07 3 20 Technical Description fig 5 1 Equipment components...
Page 122: ...20 20 Chapter 5...
Page 142: ...20 20 Chapter 6...
Page 161: ...350004900 Rev 07 5 10 Soft cushion Set fig 9 6 Use of lumbar support cushion...
Page 184: ...10 10 Chapter 11...
Page 232: ...4 4 Chapter 17...
Page 236: ...4 4 Chapter 18...