Electromagnetic interference immunity CARDIO LINE 400/400 MED
Products made by ERGO-FIT are designed for use in the environments listed below.
Please make sure to use the product in the appropriate environment.
Inerference immunity
test
IEC 60601 test
level
Compliance
level
Electromagnetic
environment
Discharge of static
electricity in comliance
with IEC 61000-4-2
+/- 6 kV contact
discharge
+/- 8 kV contact
discharge
6 kV
8 kV
Floors should be made
of wood, concrete or
laid with ceramic tiles.
If the floor is laid with
synthetic materials, the
relative humidity must
be greater than 30 %.
Electrical fast transient
bursts in compliance
with IEC 61000-4-4
+/- 2 kV for mains
cable
+/- 2 kV for lead
in and lead out
2 kV
The quality of the
supply voltage should
conform to a typical
business or hospital
environment.
Surge voltages in
compliance with IEC
61000-4-5
+/- 1 kV opposed-
mode voltage
+/- 2 kV common
voltage
2 kV
The quality of the
supply voltage should
conform to a typical
business or hospital
environment
Voltage dips, shortterm
loss and fluctuations
of the supply voltage
in compliance with IEC
61000-4-11
The quality of the
supply voltage should
conform to a typical
business or hospital
environment.
Magnetic field induced
by supply frequency
(50/60 Hz) in
compliance with
IEC 61000-4-8
The magnetic fields
induced by the supply
frequency should con-
form to typical values
as they can be found
in a typical business or
hospital environment.
Appendix
Technical and optical modifications as well as misprints reserved -
© 2011
by ERGO-FIT GmbH & Co. KG
6 0