CyBio FeliX
Maintenance & Care
8.3.4
Check/replace the O-rings
The O-rings of the CyBio FeliX are constantly subjected to mechanical stress (when using
liquid handling adapters) due to contact with the pipette tips. Thus, for safety reasons and
to ensure tightness, the O-rings on the cones must be replaced at regular intervals. If a
maintenance contract is in place, the Analytik Jena GmbH service department will perform
an O-ring replacement at each visit. Otherwise, the O-rings should be replaced if there are
visible signs of wear. This can be done with the help of the enclosed tool.
Preparation:
1.
Switch off the device
(
).
2.
Take the O-rings and slide them one
after the other over the O-ring guide
pin.
3.
Take the O-ring guide pin with the O-
rings fitted, insert it into the opening
provided and press it into the open-
ing as far as it will go.
+
=