54
Rev. B 2022/02
Operation
CyBio FeliX
Moving the pipetting head (LED but-
ton version)
Only possible when the louver
1
and/or the external E-STOP* are open.
The function is used to move the Z axis in order to be able to move the pipetting head
upwards so that it can be removed or changed.
1.
Check if LED (YELLOW) is permanently on
2.
Press LED (YELLOW) (> 3 sec. –
yellow/off
).
3.
Release LED (YELLOW).
The "Move pipetting head in Z direction" mode is activated when the
LED (YELLOW)
lights up and the operating status display flashes green.
4.
Press and
hold
the LED (YELLOW) - observe the pipetting head (is moved).
The pipetting head moves to the top vertical position and then to the end position.
Stop pipetting head
1.
Release LED (YELLOW).
To reactivate, the typing sequence must be repeated.
The open external E-STOP* signals that a monitored room - outside the device - is accessi-
ble (e.g. the door of an enclosure is open).
Adjustment mode (LED button ver-
sion)
Only possible when the louver
2
and/or the external E-STOP* are open.
The Adjustment mode function is intended for software-controlled safe operation with the
louvers open, e.g., to better view the processes when determining pipetting heights in a
microplate or determining an offset starting from a specific position.
1.
Check that the lower LED/button (YELLOW) lights up.
2.
Briefly press the LED/button (YELLOW) once (< 3 sec. -
yellow/fast flashing
).
3.
Within 2 sec. keep LED/button (YELLOW) pressed for 3 sec. - (yellow off)
and then release it.
The "Adjustment mode" is activated when the LED (YELLOW)
flashes slowly
.
With the next PC command, the movements take place at a reduced speed for the Adjust-
ment mode.
Deactivate Adjustment mode
1.
Press LED (YELLOW) (or close the louver) - (yellow off).
To reactivate, the typing sequence must be repeated.
CAUTION
Dangerous situation!
Movements are executed slowly. Nevertheless, do not reach into the device when drives
are being moved.
1 Does not apply to the version without enclosure/louver (e.g. 30-5015-500-24).
2 Does not apply to the version without enclosure/louver (e.g. 30-5015-500-24).
Summary of Contents for Analytik Jena CyBio FeliX
Page 1: ...CyBio FeliX Operating instructions...
Page 3: ...Rev B 2022 02 III CyBio FeliX...
Page 4: ...IV Rev B 2022 02 CyBio FeliX...
Page 8: ...VIII Rev B 2022 02 CyBio FeliX...
Page 12: ...XII Rev B 2022 02 CyBio FeliX...
Page 14: ...XIV Rev B 2022 02 CyBio FeliX...
Page 24: ...10 Rev B 2022 02 Technical Specifications CyBio FeliX...
Page 38: ...20 Rev B 2022 02 Safety instructions CyBio FeliX...
Page 52: ...38 Rev B 2022 02 Technical description CyBio FeliX...
Page 56: ...42 Rev B 2022 02 Packing transport and storage CyBio FeliX...
Page 88: ...74 Rev B 2022 02 CyBio FeliX...
Page 98: ...84 Rev B 2022 02 Maintenance Care CyBio FeliX...
Page 100: ...86 Rev B 2022 02 Shutting Down CyBio FeliX...