Rev. B 2022/02
53
CyBio FeliX
Operation
7.4.2
Move the pipetting head in the Z direction
Move pipetting head (arrow key ver-
sion
/
)
Only possible if the louver
1
is closed.
The function is used to move the pipetting head in the Z direction so that it can be removed
or changed.
1.
Louver closed.
2.
Press
or
(> 3 sec. –
LED (green) flashes fast
).
3.
Release
the button
4.
Press
or
and
hold
– observe the pipetting head
(is moved).
Pipetting head is moving:
to the top vertical position
down to the end position
Stop pipetting head
1.
Release
or .
To reactivate, the typing sequence must be repeated.
CAUTION
Dangerous situation!
Movements are executed slowly.
Nevertheless, do not reach into the device when drives are being moved.
1 Does not apply to the version without enclosure/louver (e.g. 30-5015-500-24).
TIP
When the operating status indicator flashes green the
“Move the pipetting head in Z direction“mode is activated.
Summary of Contents for Analytik Jena CyBio FeliX
Page 1: ...CyBio FeliX Operating instructions...
Page 3: ...Rev B 2022 02 III CyBio FeliX...
Page 4: ...IV Rev B 2022 02 CyBio FeliX...
Page 8: ...VIII Rev B 2022 02 CyBio FeliX...
Page 12: ...XII Rev B 2022 02 CyBio FeliX...
Page 14: ...XIV Rev B 2022 02 CyBio FeliX...
Page 24: ...10 Rev B 2022 02 Technical Specifications CyBio FeliX...
Page 38: ...20 Rev B 2022 02 Safety instructions CyBio FeliX...
Page 52: ...38 Rev B 2022 02 Technical description CyBio FeliX...
Page 56: ...42 Rev B 2022 02 Packing transport and storage CyBio FeliX...
Page 88: ...74 Rev B 2022 02 CyBio FeliX...
Page 98: ...84 Rev B 2022 02 Maintenance Care CyBio FeliX...
Page 100: ...86 Rev B 2022 02 Shutting Down CyBio FeliX...