Maintenance & Care
CyBio FeliX
8.3.2
Service the head mount
Fig. 47:
Head mount
Note:
1.
Check guides (rails) (cleanliness, burrs, warps, etc.)
2.
Grease the rails very thinly with silicone grease.
Wipe off any excess grease.
8.3.3
Cleaning cones
For security reasons and to ensure tightness, it is important that the cones of the liquid
handling adapters are free of soiling when using these adapters. Therefore, the cones must
be cleaned in regular intervals using a dry lint-free cloth (e.g. to remove dust).
1
1 Guides
TIP
O-rings positioned on the cones must not be displaced or turned out of position!