Rev. B 2022/02
49
CyBio FeliX
Operation
7.3
Changing the pipetting head
The pipetting head can be inserted / replaced while the device is switched on or off.
Ensure sufficient acclimatization (same temperature at the place of installation of head
and device).
The seating of the head (X axis) should be in central position (can be moved manually).
The seating of the head (Z axis) should not be in its topmost position. We recommend
an offset of - 30 mm starting from the top vertical position for replacing the head.
CAUTION
Dangerous situation!
Risk of crushing as the louver
1
is opened and closed.
1 Does not apply to the version without enclosure/louver (e.g. 30-5015-500-24).
CAUTION
Dangerous situation!
Risk of crushing - the pipetting head has a mass of approx. 6 - 7 kg.
Always hold the pipetting head with both hands.
NOTE
Make sure that the cover magazine (transport protection; OL3316-11-200) was at-
tached before removing the pipetting head because the head must only be put down
when resting on this protective cover.
This is essential for preventing damage to the device!
Summary of Contents for Analytik Jena CyBio FeliX
Page 1: ...CyBio FeliX Operating instructions...
Page 3: ...Rev B 2022 02 III CyBio FeliX...
Page 4: ...IV Rev B 2022 02 CyBio FeliX...
Page 8: ...VIII Rev B 2022 02 CyBio FeliX...
Page 12: ...XII Rev B 2022 02 CyBio FeliX...
Page 14: ...XIV Rev B 2022 02 CyBio FeliX...
Page 24: ...10 Rev B 2022 02 Technical Specifications CyBio FeliX...
Page 38: ...20 Rev B 2022 02 Safety instructions CyBio FeliX...
Page 52: ...38 Rev B 2022 02 Technical description CyBio FeliX...
Page 56: ...42 Rev B 2022 02 Packing transport and storage CyBio FeliX...
Page 88: ...74 Rev B 2022 02 CyBio FeliX...
Page 98: ...84 Rev B 2022 02 Maintenance Care CyBio FeliX...
Page 100: ...86 Rev B 2022 02 Shutting Down CyBio FeliX...