Rev. B 2022/02
9
CyBio FeliX
Technical Specifications
2.2
Overview operating data/conditions
Technical operating data
Application class
Bench-top device, closed room facilities in clean condition
Permissible altitude
up to 2000 m
Protection class
I
Mains voltage
Frequency
100 – 240 V AC ±10%
50/60 Hz
see
Connection value label on
the device (label)
Protective fusing
2 fuses 5 x 20 mm
T 4 A 250 V AC, 215.004
1
1 High Breaking Capacity Fuse; Buerklin 42 G 1786
Current consumption
2 A
Overvoltage category
II
Degree of contamination
2
Interfaces
Sub-D 9 pin socket
(RS 232 IN)
Sub-D 9 pin plug
(RS 232 forwarding)
USB socket type B
RJ 45* (Ethernet)
Airborne sound emission
< 70 db (A)
Operating and storage conditions
Operation:
• Permissible ambient temperature
• Permissible relative humidity
+15 °C to +35 °C
≤
75 % at +35 °C
non-condensing
Storage and transport:
• Permissible ambient temperature
• Permissible relative humidity
-10 °C to +50 °C
≤
85 % at +30 °C
non-condensing
Floor base requirements
Stable, horizontal, dry, free from vibration
Summary of Contents for Analytik Jena CyBio FeliX
Page 1: ...CyBio FeliX Operating instructions...
Page 3: ...Rev B 2022 02 III CyBio FeliX...
Page 4: ...IV Rev B 2022 02 CyBio FeliX...
Page 8: ...VIII Rev B 2022 02 CyBio FeliX...
Page 12: ...XII Rev B 2022 02 CyBio FeliX...
Page 14: ...XIV Rev B 2022 02 CyBio FeliX...
Page 24: ...10 Rev B 2022 02 Technical Specifications CyBio FeliX...
Page 38: ...20 Rev B 2022 02 Safety instructions CyBio FeliX...
Page 52: ...38 Rev B 2022 02 Technical description CyBio FeliX...
Page 56: ...42 Rev B 2022 02 Packing transport and storage CyBio FeliX...
Page 88: ...74 Rev B 2022 02 CyBio FeliX...
Page 98: ...84 Rev B 2022 02 Maintenance Care CyBio FeliX...
Page 100: ...86 Rev B 2022 02 Shutting Down CyBio FeliX...