8.1.4
Troubleshooting when a diagnostic message is showing
The 5081-P pH/ORP transmitter continuously monitors the measurement loop (sensor and
transmitter) for problems. If a problem is detected, the transmitter displays a fault or error
message. The message appears in the temperature/output area of the main display. The
table lists each diagnostic message and the section to consult for help.
Message
Section
GLASSFAIL
GLASSWArn
rEF FAIL
rEF WArn
CALIbrAtE
tEMP HI
tEMP LO
LInE FAIL
InPUt WArn
SLOPE Err LO
SLOPE Err HI
Std Err
rOM FAIL
CPU FAIL
AdC WArn
CyCLE PWr
WrItE Err
FACt FAIL
GLASSFAIL
GLASSFAIL
is an electrode fault message. It means the glass impedance is outside the
programmed Glass Fault High (GFH) or Glass Fault Low (GFL) limit. Glass Fault High
suggests the electrode is aging or the electrode is not immersed in the process liquid.
Glass Fault Low implies that the pH sensitive glass is cracked.
GLASSFAIL
also appears if
inappropriate limits have been entered into the transmitter.
If the measurement system was previously commissioned and operating correctly,
GLASSFAIL
likely means a real problem exists. However, if the system is being started up or
if the advanced diagnostic feature is being used for the first time,
GLASSFAIL
could be
caused by a miswired sensor or by programmed limits that are not correct for the sensor.
Diagnostics and troubleshooting
Instruction Manual
55
Summary of Contents for Rosemount5081
Page 4: ......
Page 10: ...Contents vi Rosemount 5081 ...
Page 12: ...Startup procedure 2 Rosemount 5081 ...
Page 23: ...Mounting 5081 on a pipe Figure 3 2 Installation Instruction Manual 13 ...
Page 24: ...Installation 14 Rosemount 5081 ...
Page 28: ...Wiring 18 Rosemount 5081 ...
Page 50: ...Programming basics 40 Rosemount 5081 ...
Page 130: ...Diagnostics and troubleshooting 120 Rosemount 5081 ...
Page 136: ...Digital communications 126 Rosemount 5081 ...
Page 140: ...Engineering drawings 130 Rosemount 5081 ...
Page 141: ...Engineering drawings Instruction Manual 131 ...
Page 142: ...Engineering drawings 132 Rosemount 5081 ...
Page 143: ...Engineering drawings Instruction Manual 133 ...
Page 144: ...Engineering drawings 134 Rosemount 5081 ...
Page 145: ...Engineering drawings Instruction Manual 135 ...
Page 146: ...Engineering drawings 136 Rosemount 5081 ...
Page 151: ...EU Declarations of Conformity 138 Rosemount 5081 ...