CPU FAIL and ROM FAIL
CPU FAIL means that the processing unit has failed internal tests. ROM FAIL means that the
internal memory has failed.
1.
Cycle the power. Leave the transmitter off for at least 30 seconds before returning
power to it.
If cycling the power fails to clear the error message, the CPU board probably needs
replacing.
2.
Call the factory for assistance.
AdC Error
AdC Error means the analog to digital converter has failed.
1.
Verify that the sensor wiring is correct and connections are tight. Be sure to check
connections at the junction box if one is being used.
2.
Disconnect sensor(s) and simulate temperature and sensor input.
If the transmitter does not respond to simulated signals, the analog PCB has
probably failed.
3.
Call the factory for assistance.
8.3
5081-A Diagnostics
8.3.1
Diagnostics overview
The 5081 transmitter can display diagnostic information that is useful in troubleshooting.
The diagnostics available depend on the measurement being made. To read diagnostic
information, go to the main display and press
DIAG
on the infrared remote controller. Press
NEXT
until the mnemonic for the desired information appears. Refer to the approprieate
section below for more information.
8.3.2
Diagnostic messages for dissolved oxygen
TYPE O2
Transmitter is measuring oxygen. Press
NEXT
to view diagnostics.
SEnSor Cur
Press
ENTER
to display raw current from sensor.
SEnSitvtY
Press
ENTER
to display sensitivity. Sensitivity is calculated during calibration. It is
the measured current divided by the concentration.
O CurrEnt
Press
ENTER
to display the zero current measured during calibration (note units).
bAr PreSS
Press
ENTER
to display the barometric pressure used by the transmitter during air
calibration.
5081-A-Ht
This is the model number. Press
ENTER
to display the software revision (SFtr) lev-
el. Press
NEXT
to show the hardware revision (HArdr) level.
Diagnostics and troubleshooting
92
Rosemount 5081
Summary of Contents for Rosemount5081
Page 4: ......
Page 10: ...Contents vi Rosemount 5081 ...
Page 12: ...Startup procedure 2 Rosemount 5081 ...
Page 23: ...Mounting 5081 on a pipe Figure 3 2 Installation Instruction Manual 13 ...
Page 24: ...Installation 14 Rosemount 5081 ...
Page 28: ...Wiring 18 Rosemount 5081 ...
Page 50: ...Programming basics 40 Rosemount 5081 ...
Page 130: ...Diagnostics and troubleshooting 120 Rosemount 5081 ...
Page 136: ...Digital communications 126 Rosemount 5081 ...
Page 140: ...Engineering drawings 130 Rosemount 5081 ...
Page 141: ...Engineering drawings Instruction Manual 131 ...
Page 142: ...Engineering drawings 132 Rosemount 5081 ...
Page 143: ...Engineering drawings Instruction Manual 133 ...
Page 144: ...Engineering drawings 134 Rosemount 5081 ...
Page 145: ...Engineering drawings Instruction Manual 135 ...
Page 146: ...Engineering drawings 136 Rosemount 5081 ...
Page 151: ...EU Declarations of Conformity 138 Rosemount 5081 ...