Dissolved oxygen
(continued)
Table 2-7:
Calibration
Air calibration (user must enter barometric pressure) or
calibration against a standard instrument
Free chlorine
Table 2-8:
Measurement range
0-20 ppm (mg/L) as Cl
2
Resolution
0.001 ppm (auto-ranges at 0.999 to 1.00 and 9.99 to 10.0)
Temperature correction for membrane permeability
Automatic between 0 and 50 °C (can be disabled)
pH correction
Automatic between pH 6.0 and 9.5. Manual pH correction
is also available.
Calibration
Against grab sample analyzed using portable test kit.
Total chlorine
Table 2-9:
Measurement range
0-20 ppm (mg/L) as Cl
2
Resolution
0.001 ppm (auto-ranges at 0.999 to 1.00 and 9.99 to 10.0)
Temperature correction for membrane permeability
Automatic between 5 and 35 °C (can be disabled)
Calibration
Against grab sample analyzed using portable test kit.
Ozone
Table 2-10:
Measurement range
0-10 ppm (mg/L)
Resolution
0.001 ppm (auto-ranges at 0.999 to 1.00 and 9.99 to 10.0)
Temperature correction for membrane permeability
Automatic between 41 and 95 °F (5 and 35 °C) (can be dis-
abled)
Calibration
Against grab sample analyzed using portable test kit.
Percent oxygen in gas
Table 2-11:
Measurement range
0 - 25% oxygen
Resolution
0.1% - TBD
Calibration
Air calibration (automatic measurement of barometric
pressure with internal pressure sensor)
Sample pressure
0 to 50 PSIG
Sample temperature
32 to 110 °F (0 to 43 °C)
Description and specifications
Instruction Manual
9
Summary of Contents for Rosemount5081
Page 4: ......
Page 10: ...Contents vi Rosemount 5081 ...
Page 12: ...Startup procedure 2 Rosemount 5081 ...
Page 23: ...Mounting 5081 on a pipe Figure 3 2 Installation Instruction Manual 13 ...
Page 24: ...Installation 14 Rosemount 5081 ...
Page 28: ...Wiring 18 Rosemount 5081 ...
Page 50: ...Programming basics 40 Rosemount 5081 ...
Page 130: ...Diagnostics and troubleshooting 120 Rosemount 5081 ...
Page 136: ...Digital communications 126 Rosemount 5081 ...
Page 140: ...Engineering drawings 130 Rosemount 5081 ...
Page 141: ...Engineering drawings Instruction Manual 131 ...
Page 142: ...Engineering drawings 132 Rosemount 5081 ...
Page 143: ...Engineering drawings Instruction Manual 133 ...
Page 144: ...Engineering drawings 134 Rosemount 5081 ...
Page 145: ...Engineering drawings Instruction Manual 135 ...
Page 146: ...Engineering drawings 136 Rosemount 5081 ...
Page 151: ...EU Declarations of Conformity 138 Rosemount 5081 ...