6.
Press
NEXT
.
The
ShoW Fit
submenu appears.
7.
Press
ENTER
.
The most recent fault message appears in the display.
8.
Press
NEXT
repeatedly to scroll through the stored messages.
The transmitter only remembers the three most recent messages.
nonE
appears if
there are no faults. Press
EXIT
to clear all the stored messages and return the
transmitter to the
ShoW Fit
display. If the transmitter loses power, all stored
warnings and fault messages are lost.
9.
Press
EXIT
to return to the process display.
8.1.7
Testing the transmitter by simulating the pH
Overview of simulating a pH input
This section describes how to simulate a pH input into the 5081-P pH/ORP transmitter. pH
is directly proportionsl to voltage. To simulate the pH measurement, connect a standard
millivolt source to the transmitter. If the transmitter is working properly, it will accurately
measure the input voltage and convert it to pH. Although the general procedure is the
same, the wiring details depend on the location of the preamplifier. Consult the table to
find the correct procedure.
Preamplifier located
in
Section
Transmitter
pH simulation when the preamplifier is located in the transmitter
Remote junction box
Sensor-mounted junc-
tion box
Sensor (model 381+
only)
pH simulation with 381+ sensor
Sensor (all other mod-
els)
pH simulation when preamplifier is in sensor
pH simulation when the preamplifier is located in the
transmitter
1.
Program PAMP to
transmitter
.
2.
Turn off sensor diagnostics.
3.
Turn off automatic temperature compensation. Set manual temperature
compensation to 25 °C (77 °F).
Diagnostics and troubleshooting
Instruction Manual
75
Summary of Contents for Rosemount5081
Page 4: ......
Page 10: ...Contents vi Rosemount 5081 ...
Page 12: ...Startup procedure 2 Rosemount 5081 ...
Page 23: ...Mounting 5081 on a pipe Figure 3 2 Installation Instruction Manual 13 ...
Page 24: ...Installation 14 Rosemount 5081 ...
Page 28: ...Wiring 18 Rosemount 5081 ...
Page 50: ...Programming basics 40 Rosemount 5081 ...
Page 130: ...Diagnostics and troubleshooting 120 Rosemount 5081 ...
Page 136: ...Digital communications 126 Rosemount 5081 ...
Page 140: ...Engineering drawings 130 Rosemount 5081 ...
Page 141: ...Engineering drawings Instruction Manual 131 ...
Page 142: ...Engineering drawings 132 Rosemount 5081 ...
Page 143: ...Engineering drawings Instruction Manual 133 ...
Page 144: ...Engineering drawings 134 Rosemount 5081 ...
Page 145: ...Engineering drawings Instruction Manual 135 ...
Page 146: ...Engineering drawings 136 Rosemount 5081 ...
Page 151: ...EU Declarations of Conformity 138 Rosemount 5081 ...