
•
If the message remains, go to Step 4.
4.
Confirm that the problem is with the remote preamplifier. Wire the pH sensor
directly to the transmitter. Change the menu from
PAMP=SnSr
to
trAnS
for the test
and return it to
SnSr
afterwards.
•
If the error message clears, the remote preamplifier is faulty. Replace the
preamplifier.
•
If the error message remians, the sensor has failed. Replace the sensor.
SLOPE Err LO
SLOPE Err LO
means that a two-point buffer calibration attempt has failed. The slope is too
low (<40 mV/pH) for a good measurement.
Troubleshooting SLOPE Err LO
1.
Repeat the calibration.
a. Inaccurate buffers can cause a low slope. Repeat the calibration using fresh
buffers. Alkaline buffers, pH 10 or greater, are particularly susceptible to
changing value in air or with age. If a high pH buffer was used in the failed
calibration, try a lower pH buffer when repeating the calibration. For example,
use pH 4 and 7 buffer instead of pH 7 and 10 buffer.
b. Allow adequate time for readings in buffer to become constant. If the sensor was
in a process substantially colder or hotter than the buffer, allow at least 20
minutes for readings in the buffer to stabilize. Alternatively, place the sensor in a
container of water at ambient temperatures for 20 minutes before starting the
calibration.
c. Be sure the correct buffer values are being entered during calibration.
•
If the second calibration was successful, an error was made during the first
attempt.
•
If the second calibration fails, go to Step 2.
2.
Check wiring. Connections to TB-10, TB-7, and TB-8 are particularly important.
Recalibrate the sensor.
•
if the wiring was the only problem, the sensor should calibrate.
•
If the message persists, go to Step 3.
3.
Inspect and clean the senor. Recalibrate the sensor.
•
If the sensor was dirty, it should calibrate after cleaning.
•
If the message persists, go to Step 4.
4.
Check for a faulty sensor.
•
If a spare sensor is available, connect it to the transmitter. Use the auto
calibration procedure to calibrate the sensor.
-
If the new sensor cannot be calibrated, the transmitter is faulty. Go to Step 5.
-
If the new sensor can be calibrated, the old sensor has failed.
•
If a spare sensor is not available, measure the glass impendance (
GIMP
). See
.
Diagnostics and troubleshooting
Instruction Manual
65
Summary of Contents for Rosemount5081
Page 4: ......
Page 10: ...Contents vi Rosemount 5081 ...
Page 12: ...Startup procedure 2 Rosemount 5081 ...
Page 23: ...Mounting 5081 on a pipe Figure 3 2 Installation Instruction Manual 13 ...
Page 24: ...Installation 14 Rosemount 5081 ...
Page 28: ...Wiring 18 Rosemount 5081 ...
Page 50: ...Programming basics 40 Rosemount 5081 ...
Page 130: ...Diagnostics and troubleshooting 120 Rosemount 5081 ...
Page 136: ...Digital communications 126 Rosemount 5081 ...
Page 140: ...Engineering drawings 130 Rosemount 5081 ...
Page 141: ...Engineering drawings Instruction Manual 131 ...
Page 142: ...Engineering drawings 132 Rosemount 5081 ...
Page 143: ...Engineering drawings Instruction Manual 133 ...
Page 144: ...Engineering drawings 134 Rosemount 5081 ...
Page 145: ...Engineering drawings Instruction Manual 135 ...
Page 146: ...Engineering drawings 136 Rosemount 5081 ...
Page 151: ...EU Declarations of Conformity 138 Rosemount 5081 ...