7.1.5
Standardization for pH
Standardization overview
Standardization is used to match the transmitter values with that of another transmitter.
The difference between the two pH values is converted into an equivalent voltage, called
the reference offset.
Note
If a sensor that has been calibrated with buffers is standardized, then placing it back in the buffer will
show a different measured pH than that of the buffer due to the standardization offset.
Standardization procedure
1.
Enter the CALIBRATE menu by pressing CAL on the IRC.
The CALIBRATE submenu appears.
2.
At the CALIBRATE submenu, press
NEXT
.
The STD submenu appears.
3.
Press
ENTER
to enter the STD prompt.
The process pH and temperature should be relatively stable.
4.
Take a grab sample from the process stream as close to the pH sensor as possible.
This transmitter reading is called the pH_trans.
5.
Measure the pH of the sample (pH_std) using the second pH meter.
6.
Note the current process reading (pH_curr). Calculate the corrected reading from
the equation: ph_corr = p (pH_std - pH_trans)
•
pH_corr = corrected pH value
•
pH_curr = current pH value
•
pH_std = standard instrument's pH value
•
pH_trans = measured sample's pH value
7.
Use the editing keys to enter the pH_corr value calculated using the equation above.
Press
ENTER
to save the value.
If the corrected value is acceptable, then the display shows the slope (current
electrode slope).
8.
If the slope is acceptable, press
EXIT
. If the slope is unacceptable, use the editing
keys to change it. Then press
ENTER
.
9.
To leave the CALIBRATE menu, press
EXIT
.
Measurements
44
Rosemount 5081
Summary of Contents for Rosemount5081
Page 4: ......
Page 10: ...Contents vi Rosemount 5081 ...
Page 12: ...Startup procedure 2 Rosemount 5081 ...
Page 23: ...Mounting 5081 on a pipe Figure 3 2 Installation Instruction Manual 13 ...
Page 24: ...Installation 14 Rosemount 5081 ...
Page 28: ...Wiring 18 Rosemount 5081 ...
Page 50: ...Programming basics 40 Rosemount 5081 ...
Page 130: ...Diagnostics and troubleshooting 120 Rosemount 5081 ...
Page 136: ...Digital communications 126 Rosemount 5081 ...
Page 140: ...Engineering drawings 130 Rosemount 5081 ...
Page 141: ...Engineering drawings Instruction Manual 131 ...
Page 142: ...Engineering drawings 132 Rosemount 5081 ...
Page 143: ...Engineering drawings Instruction Manual 133 ...
Page 144: ...Engineering drawings 134 Rosemount 5081 ...
Page 145: ...Engineering drawings Instruction Manual 135 ...
Page 146: ...Engineering drawings 136 Rosemount 5081 ...
Page 151: ...EU Declarations of Conformity 138 Rosemount 5081 ...