7.2.3
Standardization
Preparation of ORP standard solutions
ASTM d 1498-93 gives procedures for making iron (ii) - iron (iii) and quinhydrone ORP
standards. The iron (ii) - iron (iii) standard is recommended. It is fairly easy to make and has
a shelf life of approximately one year; in contrast, quinhydrone standards contain toxic
quinhydrone and have only an eight hour shelf life. Iron (ii) - iron (iii) standard is available
from Rosemount as PN R508-16OZ. The ORP of the standard solution measured against a
silver-silver chloride reference electrode is 476 ± 20 mV at 25 °C (77 °F).
Standardizing with an ORP sensor
1.
Press
CAL
on the IRC to enter the CALIBRATE menu.
The STD submenu appears.
2.
Rinse the sensor with DI water and place it in the ORP standard with a reference
thermometer. Ensure that the ORP sensor is submerged at least 3 in. below the
water level.
3.
Once the temperature and ORP readings are stable, press
ENTER
.
4.
Use the editing keys to change the flashing display to the desired ORP reading. Press
ENTER
to save.
5.
Press
EXIT
to return to the main display.
7.2.4
Matching transmitter to ORP sensor's temperature
element
1.
Press
PROG
on the IRC.
2.
Press
NEXT
until the TEMP submenu appears in the display. Press
ENTER
.
3.
Press
EXIT
to return to the main display.
7.3
Calibrating contacting and toroidal
conductivity sensors
7.3.1
Overview of calibrating contacting conductivity sensors
Contacting conductivity sensors do not require frequent calibrations. The sensors are
ready to be installed right out of the box.
7.3.2
Calibrating with contacting and conductivity sensors
1.
Pres
CAL
on the IRC. Press
ENTER
.
Measurements
46
Rosemount 5081
Summary of Contents for Rosemount5081
Page 4: ......
Page 10: ...Contents vi Rosemount 5081 ...
Page 12: ...Startup procedure 2 Rosemount 5081 ...
Page 23: ...Mounting 5081 on a pipe Figure 3 2 Installation Instruction Manual 13 ...
Page 24: ...Installation 14 Rosemount 5081 ...
Page 28: ...Wiring 18 Rosemount 5081 ...
Page 50: ...Programming basics 40 Rosemount 5081 ...
Page 130: ...Diagnostics and troubleshooting 120 Rosemount 5081 ...
Page 136: ...Digital communications 126 Rosemount 5081 ...
Page 140: ...Engineering drawings 130 Rosemount 5081 ...
Page 141: ...Engineering drawings Instruction Manual 131 ...
Page 142: ...Engineering drawings 132 Rosemount 5081 ...
Page 143: ...Engineering drawings Instruction Manual 133 ...
Page 144: ...Engineering drawings 134 Rosemount 5081 ...
Page 145: ...Engineering drawings Instruction Manual 135 ...
Page 146: ...Engineering drawings 136 Rosemount 5081 ...
Page 151: ...EU Declarations of Conformity 138 Rosemount 5081 ...