9.
Use the arrow keys to change the setting.
The range is 0 to 225.
10.
Press
ENTER
to save.
The screen displays the TEST prompt.
11.
Use the arrow keys to enter the desired test current. Press
ENTER
to start the test.
Press
EXIT
to end the test.
12.
Press
RESET
to return to the main display.
6.3
Temperature correction
1.
Press
PROG
. Press
NEXT
until the TEMP submenu appears on the display.
2.
Press
ENTER
.
The screen displays the
T AUTO
prompt.
3.
Use the up or down arrow keys to enable (ON) or disable (OFF) the automatic
temperature correction feature. Press
ENTER
to save.
The
T MAN
prompt is displayed.
Programming - basics
Instruction Manual
35
Summary of Contents for Rosemount5081
Page 4: ......
Page 10: ...Contents vi Rosemount 5081 ...
Page 12: ...Startup procedure 2 Rosemount 5081 ...
Page 23: ...Mounting 5081 on a pipe Figure 3 2 Installation Instruction Manual 13 ...
Page 24: ...Installation 14 Rosemount 5081 ...
Page 28: ...Wiring 18 Rosemount 5081 ...
Page 50: ...Programming basics 40 Rosemount 5081 ...
Page 130: ...Diagnostics and troubleshooting 120 Rosemount 5081 ...
Page 136: ...Digital communications 126 Rosemount 5081 ...
Page 140: ...Engineering drawings 130 Rosemount 5081 ...
Page 141: ...Engineering drawings Instruction Manual 131 ...
Page 142: ...Engineering drawings 132 Rosemount 5081 ...
Page 143: ...Engineering drawings Instruction Manual 133 ...
Page 144: ...Engineering drawings 134 Rosemount 5081 ...
Page 145: ...Engineering drawings Instruction Manual 135 ...
Page 146: ...Engineering drawings 136 Rosemount 5081 ...
Page 151: ...EU Declarations of Conformity 138 Rosemount 5081 ...