4.
Disconnect the sensor and wire transmitter as shown in
.
5.
Attach a jumper between TB-7 (REF IN) and TB-10 (pH IN).
6.
Measure the voltage. Press
DIAG
on the IRC.
The InPut voltage in millivolts appears in the temperature-output area. The main
display continues to show pH. The measured voltage should be 0 mV, and the pH
should be approximately 7. Because the calibration data in the transmitter may be
offsetting the input voltage, the displayed pH may not be exactly 7.0. If the actual
readings are close to expected, the transmitter is probably operating normally.
7.
If a standard millivolt source is available, remove the jumper between TB-7 and
TB-10 and connect the voltage source.
8.
Calibrate the transmitter. Use 0.0 mV for pH 7 (
bF1
) and -177.4 mV for pH 10 (
bF2
).
If the transmitter is working, it should accept the calibration.
9.
To check linearity, leave autocalibration and return to the main display. Set the
voltage source to the values in the table and verify that the pH reading matches the
expected value.
Voltage (mV)
pH
295.8
2.00
118.3
5.00
-118.3
9.00
-295.8
12.00
Diagnostics and troubleshooting
76
Rosemount 5081
Summary of Contents for Rosemount5081
Page 4: ......
Page 10: ...Contents vi Rosemount 5081 ...
Page 12: ...Startup procedure 2 Rosemount 5081 ...
Page 23: ...Mounting 5081 on a pipe Figure 3 2 Installation Instruction Manual 13 ...
Page 24: ...Installation 14 Rosemount 5081 ...
Page 28: ...Wiring 18 Rosemount 5081 ...
Page 50: ...Programming basics 40 Rosemount 5081 ...
Page 130: ...Diagnostics and troubleshooting 120 Rosemount 5081 ...
Page 136: ...Digital communications 126 Rosemount 5081 ...
Page 140: ...Engineering drawings 130 Rosemount 5081 ...
Page 141: ...Engineering drawings Instruction Manual 131 ...
Page 142: ...Engineering drawings 132 Rosemount 5081 ...
Page 143: ...Engineering drawings Instruction Manual 133 ...
Page 144: ...Engineering drawings 134 Rosemount 5081 ...
Page 145: ...Engineering drawings Instruction Manual 135 ...
Page 146: ...Engineering drawings 136 Rosemount 5081 ...
Page 151: ...EU Declarations of Conformity 138 Rosemount 5081 ...