3
Installation
3.1
Unpacking and inspection
Complete the following steps when you unpack your instrument.
1.
Inspect the shipping container. If there is damage, contact the shipper immediately
for instructions.
2.
If there is no apparent damage, remove the transmitter.
3.
Ensure that all items shown on the packing list are present. If items are missing,
contact your local Customer Care representative.
4.
Save the shipping container and packaging.
They can be reused to return the transmitter to the factory in case of damage.
3.2
Installation
3.2.1
General information
1.
The transmitter tolerates harsh environments. For best results, install the
transmitter in an area where temperature extremes, vibrations, and
electromagnetic and radio frequency interference are minimized or absent.
2.
To prevent unintentional exposure of the transmitter circuitry to the plant
environment, keep the security lock in place over the circuit end cap. To remove the
circuit end cap, loosen the lock nut until the tab disengages from the cap end and
then unscrew the cover.
3.
The transmitter has two 3/4 inch conduit openings, one on each side of the housing.
Run sensor cable through the left side opening (as viewed from the wiring terminal
end of the transmitter) and run power wiring through the right side opening.
4.
Use water tight cable glands to keep moisture out of the transmitter.
5.
If conduit is used, plug and seal the connections at the transmitter housing to
prevent moisture from getting inside the transmitter.
CAUTION!
EQUIPMENT DAMAGE
Moisture accumulating in the transmitter housing can affect the performance of the
transmitter and may void the warranty.
6.
If the transmitter is installed some distance from the sensor, a remote junction box
with preamplifier in the junction box or in the sensor may be necessary. Consult the
sensor instruction manual for maximum cable lengths.
Installation
Instruction Manual
11
Summary of Contents for Rosemount5081
Page 4: ......
Page 10: ...Contents vi Rosemount 5081 ...
Page 12: ...Startup procedure 2 Rosemount 5081 ...
Page 23: ...Mounting 5081 on a pipe Figure 3 2 Installation Instruction Manual 13 ...
Page 24: ...Installation 14 Rosemount 5081 ...
Page 28: ...Wiring 18 Rosemount 5081 ...
Page 50: ...Programming basics 40 Rosemount 5081 ...
Page 130: ...Diagnostics and troubleshooting 120 Rosemount 5081 ...
Page 136: ...Digital communications 126 Rosemount 5081 ...
Page 140: ...Engineering drawings 130 Rosemount 5081 ...
Page 141: ...Engineering drawings Instruction Manual 131 ...
Page 142: ...Engineering drawings 132 Rosemount 5081 ...
Page 143: ...Engineering drawings Instruction Manual 133 ...
Page 144: ...Engineering drawings 134 Rosemount 5081 ...
Page 145: ...Engineering drawings Instruction Manual 135 ...
Page 146: ...Engineering drawings 136 Rosemount 5081 ...
Page 151: ...EU Declarations of Conformity 138 Rosemount 5081 ...