Troubleshooting WritE Err
Check the position of jumper JP1 on the CPU board. If the jumper is hanging off one of the
pins, place it across both pins. If the jumper is missing entirely, use jumper JP3 (50/60 Hz),
which is not a critical jumper.
There are similar numbered jumpers on the analog board. The
jumper to be checked is on the CPU board, which is the center board in the stack.
Turn the
power to the transmitter off and then back on.
•
Toggling the power should cause the message to disappear.
•
If the message does not disappear, replace the electronic board stack.
FACt FAIL error message
FACt FAIL means the unit has not been factory calibrated. Call the factory. The transmitter
will probably need to be returned to the factory for calibration.
8.1.5
Troubleshooting when no diagnostic message is
showing
If no diagnostic message is showing, locate the symptom(s) in the table below and refer to
the appropriate section for assistance.
Symptom
Section
Id000
appears in display when trying to program
or calibrate transmitter
Error message flashing in display
Transmitter does not respond to remote con-
troller
Transmitter not responding to IRC
Calibration problems
SLOPE Err HI
or
SLOPE Err LO
appears after calibra-
tion attempt
SLOPE Err LO or SLOPE Err Hi appear after calibra-
tion attempt
bF1
or
bF2
continuously flashes during auto cali-
bration
bF1 or bF2 continuously flashes during auto cali-
bration
pH reading in buffer drifts during manual cali-
bration
pH reading in buffer drifts during manual calibra-
tion
Measurement problems
Sensor does not respond to known pH changes
Sensor not responding to pH changes
Buffer calibration is acceptable; process pH is
slightly different from expected value
Buffer calibration is acceptable; process pH is
slightly different from expected value
Buffer calibration is acceptable; process pH is
grossly wrong and/or readings are noisy
Buffer calibration is acceptable; process pH is
grossly different from expected value
Temperature reading is inaccurate
Temperature reading is inaccurate
Transmitter problems
No display
Diagnostics and troubleshooting
68
Rosemount 5081
Summary of Contents for Rosemount5081
Page 4: ......
Page 10: ...Contents vi Rosemount 5081 ...
Page 12: ...Startup procedure 2 Rosemount 5081 ...
Page 23: ...Mounting 5081 on a pipe Figure 3 2 Installation Instruction Manual 13 ...
Page 24: ...Installation 14 Rosemount 5081 ...
Page 28: ...Wiring 18 Rosemount 5081 ...
Page 50: ...Programming basics 40 Rosemount 5081 ...
Page 130: ...Diagnostics and troubleshooting 120 Rosemount 5081 ...
Page 136: ...Digital communications 126 Rosemount 5081 ...
Page 140: ...Engineering drawings 130 Rosemount 5081 ...
Page 141: ...Engineering drawings Instruction Manual 131 ...
Page 142: ...Engineering drawings 132 Rosemount 5081 ...
Page 143: ...Engineering drawings Instruction Manual 133 ...
Page 144: ...Engineering drawings 134 Rosemount 5081 ...
Page 145: ...Engineering drawings Instruction Manual 135 ...
Page 146: ...Engineering drawings 136 Rosemount 5081 ...
Page 151: ...EU Declarations of Conformity 138 Rosemount 5081 ...