Certification conditions
Connect per control drawings 1400676 or
1400677 for Intrinsically Safe Installations.
Non incendive certification
Class I, Division 2, Groups A, B, C, D
T4 T
amb
max. 70 °C
4X enclosure type
Certification conditions
For non-incendive field wiring, see installation
drawings 1400676 or 1400677.
Certification standards
Certified to FM Approvals standards: 3600:2011,
3610:2015, 3611:2016, 3615:2006, 3810:2005,
and ANSI/NEMA 250:1991
2.4
Functional specifications
The tables below display the specifications for measuring different substances with the
5081.
pH/ORP
Table 2-4:
pH range
0 to 14
ORP range
-1400 to +1400 mV
Calibration/standardization
The automatic buffer recognition uses stored buffer values and their temperature
curves for the most common buffer standards available worldwide. The transmit-
ter also performs a stabilization check on the sensor in each buffer. A manual two-
point calibration is made by immersing the sensor in two different buffer solutions
and entering the pH values. The microprocessor automatically calculates the slope
which is used for self-diagnostics. An error message is displayed if the pH sensor is
faulty. This slope can be read on the display and/or manually adjusted if desired.
An on-line, one-point process standardization is accomplished by entering the pH
or ORP value of a grab sample as measured by a lab reference.
Preamplifier location
A preamplifier must be used to convert the high impedance pH electrode signal to
a low impedance signal for transmitter use. The integral preamplifer of the 5081
may be used when the sensor to transmitter distance is less than 15 ft. (4.5 m). A
sensor with a built-in preamplifier or a junction box can be used if distance is lon-
ger than 15 ft. (4.5 m).
Automatic temperature compen-
sation
External 3 or 4 wire Pt 100 RTD or Pt 1000 RTD, located in the sensor, compen-
sates the pH reading for temperature fluctuations. Compensation covers the
range 5 to 270 °F (-15 to 130 °C). Manual temperature compensation is also select-
able.
Accuracy
±01 mv at 25 °C (77 °F) ± 0.01 pH
Repeatability
±01 mv at 25 °C (77 °F) ±0.01 pH
Stability
0.25% / year at 25 °C (77 °F)
Description and specifications
Instruction Manual
7
Summary of Contents for Rosemount5081
Page 4: ......
Page 10: ...Contents vi Rosemount 5081 ...
Page 12: ...Startup procedure 2 Rosemount 5081 ...
Page 23: ...Mounting 5081 on a pipe Figure 3 2 Installation Instruction Manual 13 ...
Page 24: ...Installation 14 Rosemount 5081 ...
Page 28: ...Wiring 18 Rosemount 5081 ...
Page 50: ...Programming basics 40 Rosemount 5081 ...
Page 130: ...Diagnostics and troubleshooting 120 Rosemount 5081 ...
Page 136: ...Digital communications 126 Rosemount 5081 ...
Page 140: ...Engineering drawings 130 Rosemount 5081 ...
Page 141: ...Engineering drawings Instruction Manual 131 ...
Page 142: ...Engineering drawings 132 Rosemount 5081 ...
Page 143: ...Engineering drawings Instruction Manual 133 ...
Page 144: ...Engineering drawings 134 Rosemount 5081 ...
Page 145: ...Engineering drawings Instruction Manual 135 ...
Page 146: ...Engineering drawings 136 Rosemount 5081 ...
Page 151: ...EU Declarations of Conformity 138 Rosemount 5081 ...