CALIbrAtE
CALIbrAte
is a diagnostic intended for future use. If the
CALIbrAte
message is showing,
disable
CALIbrAte
.
tEMP HI and tEMP LO
tEMP HI and tEMP LO mean the transmitter has detected a problem with the temperature
measuring circuit. The problem may lie in the sensor, the cable, or the transmitter. The
determination of temperature is an integral part of the pH measurement. Therefore,
failure of the temperature measuring circuit is a system disabling condition. However, in
an emergency, automatic temperature compensation can be disabled and the transmitter
placed in manual temperature compensation. For manual temperature compensation,
choose a temperature equal to the average temperature of the process. The resulting pH
reading will be in error. The more variable the temperature and the further the pH from 7,
the greater the error.
Troubleshooting flowchart - tEMP HI and tEMP LO
1.
Check wiring, jumper settings, and software settings.
a. Check the wiring between the sensor and the transmitter. Pay particular
attention to TB=3 (RTD RTN), TB-4 (RTD SN), and TB-5 (RTD RTN ).
NOTICE
TB-3 means terminal 3 on the terminal board.
b. Be sure the software settings match the type of RTD in the sensor.
•
If the diagnostic message disappears, the sensor is in good condition.
•
If the message persists, go to Step 2.
2.
Check the sensor. Disconnect the RTD leads and measure the resistances shown in
. The measured resistance should agree with the value in
within about 1%. If the measured resistance is appreciably different (between 1 and
5%) from the value shown, the discrepancy can be calibrated out.
Diagnostics and troubleshooting
Instruction Manual
61
Summary of Contents for Rosemount5081
Page 4: ......
Page 10: ...Contents vi Rosemount 5081 ...
Page 12: ...Startup procedure 2 Rosemount 5081 ...
Page 23: ...Mounting 5081 on a pipe Figure 3 2 Installation Instruction Manual 13 ...
Page 24: ...Installation 14 Rosemount 5081 ...
Page 28: ...Wiring 18 Rosemount 5081 ...
Page 50: ...Programming basics 40 Rosemount 5081 ...
Page 130: ...Diagnostics and troubleshooting 120 Rosemount 5081 ...
Page 136: ...Digital communications 126 Rosemount 5081 ...
Page 140: ...Engineering drawings 130 Rosemount 5081 ...
Page 141: ...Engineering drawings Instruction Manual 131 ...
Page 142: ...Engineering drawings 132 Rosemount 5081 ...
Page 143: ...Engineering drawings Instruction Manual 133 ...
Page 144: ...Engineering drawings 134 Rosemount 5081 ...
Page 145: ...Engineering drawings Instruction Manual 135 ...
Page 146: ...Engineering drawings 136 Rosemount 5081 ...
Page 151: ...EU Declarations of Conformity 138 Rosemount 5081 ...