120
Reference Manual
00809-0100-4840, Rev CB
Appendix F: Programming the 3102 and 3105 using HART
February 2015
Programming the 3102 and 3105 using HART
F.2.2
Set as empty
If the bottom reference is
unknown
and the tank is
empty
, the transmitter can change the
Transmitter Bottom Reference value to the Distance measurement with the tank empty.
P010 = (D910 - P060)
Where:
P010 = Transmitter Bottom Reference setting (see
).
D910 = Distance measurement (see
).
P060 = Distance Offset setting (see
).
Field Communicator
To select the Set As Empty command
1.
From the
Home
screen, select 2: Configure.
2.
Select 2: Manual Setup.
3.
Select 2: Level.
4.
Select 3: Present Depth.
5.
Select 2: Set as Empty, and then select “ENTER”.
Note
When messages appear, take appropriate action if needed and select “OK”.
Note
The command Set As Empty is also at Fast Key sequence 2, 1, 2.
Rosemount 3490 Series Control Unit
To select the Set As Empty command
1.
From the
Main Menu
screen, select SETUP.
2.
Select the transmitter (e.g. “Tx1: 3102”).
3.
Select DUTY.
4.
Select SET AS EMPTY.
5.
Follow on-screen instructions to perform the Set As Empty action.
(If prompted to change the mode to off-line, use the Enter key).
6.
Select “
Quit
” to exit to the previous menu.
Fast Keys
2, 2, 2, 3, 2
Select:
1. Present Depth
2. Set as Empty
3. Exit
2. Set as Empty
ENTER
ABORT
(
Field Communicator Screen Shown
)
Esc=Quit =Start
SET AS EMPTY
(
Rosemount 3491 Screen Shown
)
Summary of Contents for Rosemount 3101
Page 2: ......
Page 4: ...ii Reference Manual 00809 0100 4840 Rev CB Title Page February 2015 ...
Page 12: ...x Reference Manual 00809 0100 4840 Rev CB Contents February 2015 Contents ...
Page 34: ...22 Reference Manual 00809 0100 4840 Rev CB Section 3 Installation February 2015 Installation ...
Page 68: ...56 Reference Manual 00809 0100 4840 Rev CB Section 4 Starting up February 2015 Starting up ...
Page 236: ...224 Reference Manual 00809 0100 4840 Rev CB Index February 2015 Index ...
Page 237: ......
Page 238: ......
Page 239: ......