8.
To change the fixed current value to a new value, press the blue button
to return to
the “LOOP” menu and re-start at step (
).
9.
Press the green button
to return to displaying the “tESt” menu option.
10.
From “tESt”, press the red button
to change to the PV display.
Alternatively, from “tESt”, press and hold both the blue
and the red
buttons
together for two seconds to get to the Engineering “Eng” menu
(see
“Engineering menu for the 3102 and 3105” on page 62
5.5
Engineering menu for the 3102 and 3105
Menu structures for this section are in
Appendix C: Integrated Display Menus
Note
If using a HART Master Device for programming the 3102 or 3105, refer to the following
sections for menu structures and parameters:
Appendix D: Rosemount 3490 Series Menus
Appendix E: Field Communicator Menus
Appendix F: Programming the 3102 and 3105 using HART
5.5.1
Accessing the engineering menu (on the 3102/3105)
Screen display:
Eng
The transmitter can be fine-tuned if site or application conditions are unusual. It is
recommended that all operational fine tuning parameters remain at the factory default settings
unless there is a good understanding of the function and capability of the parameters.
The “reload factory defaults” function is found in this menu, and should be used if the
transmitter has been configured incorrectly, or if the transmitter needs to be reset to factory
default values.
To reach the Engineering menu option “Eng”
1.
Begin from the PV display.
(
is a map of the menu structure).
2.
Press the blue button
to indicate the “DiAg” menu option.
(To exit to the PV display, press the red button
).
3.
Hold down the blue button
for two seconds, and then release.
The display changes to the “tEst” menu option. (To exit to the PV display, press
).
4.
Hold down both the blue button
and red button
for two seconds.
The display changes to the “Eng” menu option. (To exit to the PV display, press
).
5.
Press the green button
to enter the “Eng” menu. The display indicates the first
engineering menu option “t.hoLd” (see below).
Summary of Contents for Rosemount 3101
Page 2: ......
Page 4: ...ii Reference Manual 00809 0100 4840 Rev CB Title Page February 2015 ...
Page 12: ...x Reference Manual 00809 0100 4840 Rev CB Contents February 2015 Contents ...
Page 34: ...22 Reference Manual 00809 0100 4840 Rev CB Section 3 Installation February 2015 Installation ...
Page 68: ...56 Reference Manual 00809 0100 4840 Rev CB Section 4 Starting up February 2015 Starting up ...
Page 236: ...224 Reference Manual 00809 0100 4840 Rev CB Index February 2015 Index ...
Page 237: ......
Page 238: ......
Page 239: ......