206
Reference Manual
00809-0100-4840, Rev CB
Appendix F: Programming the 3102 and 3105 using HART
February 2015
Programming the 3102 and 3105 using HART
F.4.13
Frequency (D916)
This indicates the transmitter’s actual operating frequency.
The factory default setting is to automatically set the operating frequency for optimum
performance (see Target Frequency on
Field Communicator or AMS Device
Manager
To view the actual operating frequency
1.
From the
Home
screen, select 3: Service Tools.
2.
Select 1: Alerts.
3.
Select 2: Diagnostic Data.
4.
Select 3: Operation, and then select 1: Transducer Frequency D916.
5.
When finished, select EXIT to exit to the previous menu.
Note
This is also at the Fast Key sequence 2, 2, 7 [or 8], 6, 2
(see
Appendix E: Field Communicator Menus
for a menu tree diagram).
Rosemount 3490 Series Control Unit
To view the actual operating frequency
1.
From the
Main Menu
screen, select MONITOR.
2.
Select the transmitter (e.g. “Tx1: 3102”).
3.
Select DIAGNOSTICS, and then select OPERATION.
4.
Select Xducer Freq.
5.
When finished, use the ESC key to exit to the previous menu.
Fast Keys
3, 1, 2, 3, 1
Transducer Frequency D916
51 kHz
EXIT
HELP
(
Field Communicator Screen Shown
)
Xducer Freq D916
51kHz
(
Rosemount 3491 Screen Shown
)
Summary of Contents for Rosemount 3101
Page 2: ......
Page 4: ...ii Reference Manual 00809 0100 4840 Rev CB Title Page February 2015 ...
Page 12: ...x Reference Manual 00809 0100 4840 Rev CB Contents February 2015 Contents ...
Page 34: ...22 Reference Manual 00809 0100 4840 Rev CB Section 3 Installation February 2015 Installation ...
Page 68: ...56 Reference Manual 00809 0100 4840 Rev CB Section 4 Starting up February 2015 Starting up ...
Page 236: ...224 Reference Manual 00809 0100 4840 Rev CB Index February 2015 Index ...
Page 237: ......
Page 238: ......
Page 239: ......